Gas Sampling Bulb

size 250 mL, glass stopcock

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali


About This Item

Codice UNSPSC:
41104008
NACRES:
NB.24
Prezzi e disponibilità al momento non sono disponibili

Materiali

glass stopcock

tecniche

whole air sampling: suitable

Misura

250 mL

applicazioni

air monitoring
environmental
industrial hygiene

Cerchi prodotti simili? Visita Guida al confronto tra prodotti

Descrizione generale

Use these glass bulbs to trap and transport a gas sample, then remove an aliquot (through the plug-type septum) for analysis. Oven annealed to resist damage during transportation and on-site use. Available with PTFE or high vacuum glass stopcocks. One half-hole cylindrical Thermogreen septum is included with each bulb.

Scegli una delle versioni più recenti:

Certificati d'analisi (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documenti section.

Se ti serve aiuto, non esitare a contattarci Servizio Clienti

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Questions

  1. Good morning, Could you tell me if they have a key with an escape valve for safety? How do I prevent the glass from breaking with the pressure of N2 gas? Thank you!

    1 answer
    1. There is no key mechanism on the gas sampling bulbs. These are intended for use at atmospheric pressure. The maximum psi has not been established. The way to keep from breaking the glass is to not overfill the apparatus.
      Directions for filling a Gas Sampling Bulb and Making a Standard:

      I. Flush the bulb with clean helium or nitrogen before collecting the sample.

      a. Open both stopcocks and flow at least on 1 1/2 to 2 bulb volumes.
      b. Close the outlet stopcock, then the inlet stop cock after turning the gas supply off.
      c. If pressure cannot be controlled then try the following:
      i. Attach a 1-2 foot piece of tubing to the outlet of the sampling bulb.
      ii. Close the inlet stopcock, then very quickly close the outlet stop cock (the tubing on the outlet will help prevent diffusion of the outside air entering the sampling bulb.
      1. Sample will be very close to atmospheric pressure after turning the gas supply off.
      iii. Inject a known quantity of compound
      iv. If necessary, heat the bulb in an oven to vaporize the compounds injected.