Accéder au contenu
Merck

HPLC Analysis of Bisphenol A in Water on Ascentis® Express C18 after SPE using Supelclean ENVI-18

HPLC Analysis of Bisphenol A in Water on Ascentis® Express C18 after SPE using Supelclean™ ENVI-18 application for HPLC

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

Colonne de CLHP Ascentis® Express C18, 2,7 μm

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

Tube ou plaque pour SPE

Réf. du produit
Description
Tarif

Tube SPE Supelclean ENVI-18

glass hardware, PTFE frit, bed wt. 500 mg, volume 6 mL, pk of 30

Étalon

Réf. du produit
Description
Tarif

Bisphenol A

≥99%

CONDITIONS

SPE tube/cartridge

Supelclean ENVI-18, 500 mg/6 mL glass SPE tube (54331-U)

condition

1 mL 1% formic acid in acetonitrile, then 1 mL water

sample

5 mL spiked water blank, concentration of 0.2 μg/mL

elution

2 mL 1% formic acid in acetonitrile

eluate post-treatment

1 mL aliquot of eluent, and evaporate/reconstitute to 0.5 mL in acetonitrile

column

Ascentis Express C18, 10 cm x 2.1 mm I.D., 2.7 μm particles (53823-U)

mobile phase

[A] water; [B] acetonitrile; (60:40, A:B)

flow rate

0.4 mL/min

pressure

3270 psi (256 bar)

column temp.

35 °C

detector

UV, 230 nm

injection

1 μL

Description

Remarque sur l'analyse

Bisphenol A is a contaminant that can leach into aqueous solutions from polycarbonate plastics or metal cans; it may impact human development and reproduction. This application demonstrates the extraction and analysis of bisphenol A from water.

Informations légales

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
Supelclean is a trademark of Sigma-Aldrich Co. LLC