HPLC Analysis of Statins in Rat Plasma on Ascentis® Express C18 after SPE using Supel™-Select HLB
application for SPE, application for LC-MS
About This Item
Technique(s)
LC/MS: suitable
solid phase extraction (SPE): suitable
Paramètres du test
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: rat plasma spiked with provastin and atorvastin at 5 and 100 ng/mL
SPE tube/cartridge: Supel-Select HLB SPE, 30 mg/1 mL (54181-U)
condition: 0.5 mL methanol:acetonitrile (1:1, v/v); 0.5 mL DI water
sample addition: 0.5 mL
washing: 0.5 mL 5% methanol
elution: 0.5 mL methanol:acetonitrile (1:1, v/v)
column: Ascentis Express C18, 5 cm x 2.1 mm, 2.7 μm particles (53822-U)
eluate post-treatment: evaporate to dryness with nitrogen at 37 °C, reconstitute in 0.5 mL DI water
mobile phase: 0.1% acetic acid diluted in 60% methanol
flow rate: 0.2 mL/min
column temp.: 35 °C
detector: ABI 3200 QT; ESI(+), MRM (423.3/321.3, 423.3/101.0, 557.3/397.2 and 557.3/453.4)
injection: 5 μL
Adéquation
application for LC-MS
application for SPE
Application(s)
clinical
Informations légales
Colonne analytique
Tube ou plaque pour SPE
Choose from one of the most recent versions:
Certificats d'analyse (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Service Clients
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique