CVSC73TP3
Multimedia Durapore® 1.2/0.5/0.22 µm, Cartridge
pore size 1.2/0.5/0.22 μm, cartridge nominal length 30 in. (75 cm), Code 7 (2-226; O-rings w/locking-tabs)
Synonyme(s) :
Durapore Multimedia Cartridge Filter 30 in. 1.2+0.5+0.22 μm Code 7
About This Item
Matériaux
PVDF
mixed cellulose esters (MCE)
mixed cellulose esters (MCE) prefilter
polypropylene
polypropylene support
silicone seal
Niveau de qualité
Conformité réglementaire
meets FDA Indirect Food Additive requirements cited in 21 CFR 177-182 (all component materials)
Stérilité
non-sterile
Mode de stérilisation compatible
autoclavable compatible
steam-in-place compatible
Gamme de produits
Durapore®
Caractéristiques
hydrophilic
Fabricant/nom de marque
Durapore®
Paramètres
≤32.4 mL/min air diffusion at 2.75 bar (40 psig) and 23 °C (in water)
0.35 bar max. differential pressure (5 psid) at 135 °C (Forward)
0.35 bar max. inlet pressure (5 psi) at 135 °C (Forward)
1.7 bar max. differential pressure (25 psid) at 80 °C (Forward)
1.7 bar max. inlet pressure (25 psi) at 80 °C (Forward)
3.5 bar max. differential pressure (50 psid) at 25 °C (Reverse; intermittent)
3.5 bar max. inlet pressure (50 psi) at 25 °C (Reverse; Intermittent)
5.5 bar max. differential pressure (80 psid) at 25 °C (Forward)
5.5 bar max. inlet pressure (80 psi) at 25 °C (Forward)
Technique(s)
sterile filtration: suitable
Longueur
30 in.
Largeur
2.7 in.
Longueur nominale des cartouches
30 in. (75 cm)
Diamètre
6.9 cm (2.7 in.)
Surface de filtration
1.68 m2
Impuretés
≤0.5 EU/mL bacterial endotoxins (LAL test, Aqueous extraction)
<0.5 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
Extractibles par gravimétrie
≤210 mg/cartridge
Matrice
Multimedia Durapore®
pore size
1.2/0.5/0.22 μm pore size
Entrée
sample type liquid
Point de bulle
≥3450 mbar (50 psig), air with water at 23 °C
Code de la cartouche
Code 7 (2-226; O-rings w/locking-tabs)
Description générale
Conditionnement
Autres remarques
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer Wetting Instructions for Filter Units with Durapore® Membrane user guide
- Storage Statement: Store in dry location
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.
Notes préparatoires
6 autoclave or SIP cycles of 30 min @ 123 °C
Remarque sur l'analyse
Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
Informations légales
Vous ne trouvez pas le bon produit ?
Essayez notre Outil de sélection de produits.
Code de la classe de stockage
11 - Combustible Solids
Classe de danger pour l'eau (WGK)
WGK 2
Certificats d'analyse (COA)
Recherchez un Certificats d'analyse (COA) en saisissant le numéro de lot du produit. Les numéros de lot figurent sur l'étiquette du produit après les mots "Lot" ou "Batch".
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique