HPLC Analysis of Superoxide Dismutase on Discovery® BIO PolyMA-WAX

application for HPLC

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Technique(s)

HPLC: suitable

Paramètres du test

column: Discovery BIO PolyMA-WAX, 5 cm × 4.6 mm, 5 μm particles (59602-U)
mobile phase: ([A] 20 mM Tris, pH 8.0 with hydrocloric acid; [B] 20 mM Tris, 0.5 M sodium chloride, pH 8.0 with hydrochloric acid)
gradient: 0 to 10% B in 30 min. (linear)
flow rate: 0.5 mL/min
column temp.: 25 °C
detector: UV, 280 nm
injection: 10 μL

Adéquation

application for HPLC

Application(s)

life science and biopharma

Remarque sur l'analyse

Discovery BIO PolyMA-SCX and PolyMA-WAX columns give sharp, efficient peaks for a wide variety of proteins.

Informations légales

Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyte

Description

    Superoxide dismutase

Choose from one of the most recent versions:

Certificats d'analyse (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Service Clients

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique