The product issue you are reporting should be investigated by our Technical Service team who can assist you with further troubleshooting and/or investigation. We kindly ask you to navigate to the link https://www.sigmaaldrich.com/techservice, and click on "Report Product Issues" under the Products Section with all the required information so that a member of our team can reach out to you to assist further. Thank you.
Recommended Products
grade
analytical standard
description
refined
CofA
current certificate can be downloaded
packaging
ampule of 1000 mg
technique(s)
HPLC: suitable
gas chromatography (GC): suitable
density
0.910 g/cm3
application(s)
food and beverages
format
neat
functional group
ester
oleic acid
shipped in
ambient
storage temp.
2-30°C
Looking for similar products? Visit Product Comparison Guide
General description
Application
Biochem/physiol Actions
Storage Class Code
10 - Combustible liquids
WGK
awg
Flash Point(F)
235.4 °F - closed cup
Flash Point(C)
113 °C - closed cup
Choose from one of the most recent versions:
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Articles
Learn more about Biodiesel as a renewable, alternative diesel fuel produced from vegetable oils, animal fats, or recycled restaurant grease.
Learn more about Biodiesel as a renewable, alternative diesel fuel produced from vegetable oils, animal fats, or recycled restaurant grease.
Learn more about Biodiesel as a renewable, alternative diesel fuel produced from vegetable oils, animal fats, or recycled restaurant grease.
Learn more about Biodiesel as a renewable, alternative diesel fuel produced from vegetable oils, animal fats, or recycled restaurant grease.
Chromatograms
application for HPLCapplication for HPLC-
I was wondering, what could be the reason to see the FAME peaks in a highly different ratio? (i.e 16:0 relative abundance is larger then 18:1). The chromatogram peaks are very clear and separated, what FAME derivatization method would you suggest to use?
1 answer-
Helpful?
-
-
What capillary column can be used to analyze Product 47118, Olive Oil, after prepared into methyl esters?
1 answer-
We used an SP™-2380 capillary column, L × I.D. 30 m × 0.25 mm, df 0.20 μm (Supelco Product No. 24110-U) to analyze the methyl esters of product 47118 Olive Oil. The temperature program is 150 °C to 250 °C, increasing at a rate of 4 °C per minute, with helium carrier gas at 20 cm/sec at 150 °C. Use an injection volume of 3 μL, and split of 100:1, and FID detection at 260 °C.
Helpful?
-
-
Is Product 47118, Olive Oil, in the form of methyl esters?
1 answer-
No, Product No. 47118 is olive oil with no modifications. The methyl esters may be prepared (for analysis) by following the procedures outlined in AOAC Method 969.33 or AOCS Method Ce 2-66.
Helpful?
-
-
What is the Department of Transportation shipping information for this product?
1 answer-
Transportation information can be found in Section 14 of the product's (M)SDS.To access the shipping information for this material, use the link on the product detail page for the product.
Helpful?
-
-
How do I obtain a Certificate of Composition for Product 47118, Olive Oil?
1 answer-
The Certificate of Composition for Product No. 47118 is shipped with the product and is not available on our website. If you have misplaced this document, please email [email protected] with the lot number to obtain a copy of the Certificate of Composition.