HPLC Analysis of Carotenes on Ascentis® Express RP-Amide, Methanol Mobile Phase

application for HPLC

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Technique(s)

HPLC: suitable

Paramètres du test

column: Ascentis Express RP-Amide, 15 cm x 4.6 mm I.D., 2.7 μm particles (53931-U)
mobile phase: methanol
flow rate: 1 mL/min
pressure: 1653 psi (114 bar)
column temp.: 30 °C
detector: UV-VIS, 454 nm
injection: 5 μL
sample: V8 Splash® fruit juice was extracted with hexane, the extract dissolved 1:1 into ethanol for HPLC injection

Adéquation

application for HPLC

Application(s)

vitamins, nutraceuticals, and natural products

Remarque sur l'analyse

Analysis was conducted using juice extract, detection via UV.

Informations légales

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
V8 Splash is a registered trademark of CSC Brands, LP

Colonne analytique

Étalon

Réf. du produit
Description
Tarif

Choose from one of the most recent versions:

Certificats d'analyse (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Service Clients

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique