Accéder au contenu
Merck
Toutes les photos(1)

Documents

G005662

HPLC Analysis of Aflatoxins on Ascentis® Express C18 with Fluorescence Detection

application for HPLC

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Technique(s)

HPLC: suitable

Paramètres du test

column: Ascentis Express C18, 15 cm x 4.6 mm I.D., 2.7 μm particles (53829-U)
mobile phase: [A] water; [B] acetonitrile; (75:25, A:B)
flow rate: 2.0 mL/min
pressure: 6381 psi (440 bar)
column temp.: 35 °C
detector: fluorescence, ex 360 nm, em 440 nm
injection: 50 μL
sample: 200 μL aflatoxin standard in acetonitrile; 0.5 μg/mL B2 and G2; 2 μg/mL B1 and G1; 800 μL delivatization solution: 70% water, 20% TFA and 10% acetic acid
heat for 9 min at 65 °C

Adéquation

application for HPLC

Application(s)

food and beverages

Informations légales

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyte

Description

    Aflatoxin G1
    Aflatoxin B1
    Aflatoxin G2
    Aflatoxin B2

Choose from one of the most recent versions:

Certificats d'analyse (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Service Clients

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique