G003766
HPLC Analysis of Diazepam and Metabolites in Serum on Discovery® C18 after SPE using Discovery® DSC-8
application for SPE, application for HPLC
About This Item
Technique(s)
HPLC: suitable
solid phase extraction (SPE): suitable
Paramètres du test
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: goat serum spiked with diazepam and metabolites at 0.5 μg/mL and diluted 1:1 in 25 mM ammonium formate, pH 7.1
SPE well plate: Discovery DSC-8 96-well SPE Plate, 100 mg/well (575627-U)
condition: 1 mL methanol followed by 1 mL DI water.
sample addition: 1 mL spiked goat serum
washing: 1 mL 25% methanol in 25 mM ammonium formate, pH 2.75
elution: 1 mL 60% methanol in 25 mM ammonium formate, pH 2.75
elution: Evaporate eluate with nitrogen purge (30 °C; ~15 min.). Reconstitute in 200 μL mobile phase.
column: Discovery C18, 5 cm x 4.6 mm I.D., 5 μm particles (504947)
mobile phase: [A] Methanol; [B]10 mM ammonium acetate, pH 4.5;(45:55, A:B)
flow rate: 1.5 mL/min
column temp.: 35 °C
detector: UV, 240 nm
injection: 25 μL
Adéquation
application for HPLC
application for SPE
Application(s)
clinical
forensics and toxicology
Informations légales
Colonne analytique
Faites votre choix parmi les versions les plus récentes :
Certificats d'analyse (COA)
Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.
Si vous avez besoin d'assistance, veuillez contacter Service Clients
Déjà en possession de ce produit ?
Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.
Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..
Contacter notre Service technique