Accéder au contenu
Merck
Toutes les photos(1)

Principaux documents

80-207

Supelco

HPA dans l′eau

suitable for GC, application for SPE

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Technique(s)

gas chromatography (GC): suitable
solid phase extraction (SPE): suitable

Paramètres du test

sample preparation: SPE (Solid Phase Extraction)
sample/matrix: water spiked with PAHs
SPE tube/cartridge: Supelclean ENVI-18, 500 mg/6 mL (57064)
condition: 2 x 6 mL toluene:methanol (10:1); 6 mL methanol; 6 mL DI water
sample addition: 250 mL water sample at 10 mL/min
drying: 10 min under vacuum
elution: 2 x 1 mL toluene:methanol (10:1)
column: 5% phenyl polysiloxane, 30m × 0.25 mm I.D., 0.25 μm film (Supelco equivalent, SLB-5ms, 28471-U)
oven: 70 °C (2 min) to 280 °C at 8 °C/min
detector: FID, 310 °C
carrier gas: helium
injection: 1 μL

Adéquation

application for SPE

Application(s)

environmental
food and beverages

Informations légales

Supelclean is a trademark of Sigma-Aldrich Co. LLC

Analyte

Description

    Dibenz[a,h]anthracene
    Benz[a]anthracene
    Indeno[1,2,3-cd]pyrene

Tube ou plaque pour SPE

Réf. du produit
Description
Tarif

Faites votre choix parmi les versions les plus récentes :

Certificats d'analyse (COA)

Lot/Batch Number

Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.

Si vous avez besoin d'assistance, veuillez contacter Service Clients

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..