Accéder au contenu
Merck
Toutes les photos(1)

Documents

CZRV73TP1

Millipore

Filtre cartouche Viresolve® NFR

Retrovirus removal

Synonyme(s) :

Viresolve NFR Cartridge Filter 30 in. Code 7

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Code UNSPSC :
23151806
eCl@ss :
32031690

Matériaux

polyethersulfone
polypropylene
silicone seal

Niveau de qualité

Caractéristiques

holdup volume 490 mL

Fabricant/nom de marque

Viresolve®

Paramètres

≤48 mL/min air diffusion at 3.45 bar (50 psig) and 23 °C (in water)
1800-3600 L process volume
3.4 bar max. differential pressure (50 psid) at 25 °C (Reverse)
5.5 bar max. differential pressure (80 psid) at 25 °C (Forward)

Technique(s)

protein purification: suitable

Longueur nominale des cartouches

30 in. (75 cm)

Surface de filtration

1.281 m2

Impuretés

≤0.5 EU/mL bacterial endotoxins (LAL test, Aqueous extraction)
<0.5 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)

Extractibles par gravimétrie

≤105 mg/cartridge

Matrice

Viresolve® NFR

Code de la cartouche

Code 7

Vous recherchez des produits similaires ? Visite Guide de comparaison des produits

Description générale

Device Configuration: Cartridge

Application

Retrovirus removal

Conditionnement

Double Easy-Open bag

Autres remarques

These products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.
Directions for Use

  • Organism Retention: Retrovirus
  • Mode of Action: Filtration (size exclusion)
  • Application: Protein purification
  • Intended Use: Viral clearance
  • Instructions for Use: Please see the user guide (pg. 3) shipped with this product
  • Storage Statement: Store at room temperature
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.

Notes préparatoires

Sterilization Method
After wetting, may be autoclaved for 3 cycles of up to 60 min at 125 °C,

using liquid cycle, slow exhaust.
This product was manufactured with a Viresolve® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).

Remarque sur l'analyse

Gravimetric Extractables: after a 10 minute 1.5 LPM/ft² flush and 24 hours in water at controlled room temperature
Will meet the USP Oxidizable Substances Test requirements after a water flush of ≥12 L

Informations légales

VIRESOLVE is a registered trademark of Merck KGaA, Darmstadt, Germany

Certificats d'analyse (COA)

Recherchez un Certificats d'analyse (COA) en saisissant le numéro de lot du produit. Les numéros de lot figurent sur l'étiquette du produit après les mots "Lot" ou "Batch".

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique