Direkt zum Inhalt
Merck
HomeChemical Analysis for Food and Beverage TestingAntibiotic Analysis Tetracyclines Sulfonamides Quinolones

Analysis of Antibiotics— Tetracyclines, Sulfonamides and Quinolones Residues in Pork Meat by LC-MS acc. to GB 31658.17-2021

Jack Wang
R&D APAC lab, Shanghai, China

 

Abstract

A sensitive liquid chromatography-mass spectrometry method was developed for the determination of antibiotic drug residues in pork meat in accordance with the GB 31658.17-2021. Target analytes included 19 sulfonamides, 13 quinolones and three tetracycline drugs. Samples were extracted with Mcllvaine Na2EDTA buffer solution, purified by solid-phase extraction using a modern HLB phase (Supel™ Swift HLB), analyzed by liquid chromatography-tandem mass spectrometry employing a Purospher® STAR RP-18e UHPLC column, and quantified by the external standard method. The developed method complied with the performance criteria required by the GB 31658.17-2021.

 Section Overview

Introduction

The widespread use of antibiotics in animal husbandry and their associated residues in animal meat has been a worldwide concern, as they are known to have detrimental health effects for the consumers.1,2 Sulfonamides, quinolones, and tetracycline antibiotics are widely used in livestock and poultry breeding. There is therefore a need to monitor these residues in meat. As these veterinary drugs are often used at random, the analytical challenge is the simultaneous detection of multiple antibiotic classes within the same sample. At present, HPLC and LC-MS/MS methods are mainly used for the analysis and detection of antibiotics, whereas LC-MS/MS is the main analysis and detection technique.3,4

Many national and international authorities and limits and issued detection methods for various matrices. For China, the GB 31658.17-2021 method5 specifies the type of substrate for testing samples as: muscle, liver and kidney tissue of cattle, such as sheep, pigs, and chicken. The method describes the residue determination of 19 sulfonamides, 13 quinolones, and 3 tetracyclines in the aforementioned matrices. The samples are extracted with Mcllvaine-Na2EDTA buffer solution, then purified by solid-phase extraction (SPE) using a hydrophilic lipophilic balanced (HLB) SPE phase, and the target compounds are determined by liquid chromatography-tandem mass spectrometry (LC-MS/MS) using an external matrix-matched calibration for quantification.

As performance & quality control criteria, the GB 31658.17-2021 defines requirements as follows:

  • Limits of Detection (LOD) and Limits of Quantification (LOQ) with LOD £5 μg/kg and LOQ £15 μg/kg.
  • Recovery at the addition concentration level of 10-500 μg/kg to be in the range of 60%-110%.
  • The relative standard deviation (%RSD) of the precision test should be less than 15%.

For the here described application, the sample pre-treatment was carried out in accordance with the GB 31658.17-2021 standard using a Supel™ Swift HLB SPE tube. 19 sulfonamides, 13 quinolones, and 3 tetracyclines (Table 2) were analyzed and quantified against a matrix match calibration by LC-MS/MS using a Purospher® STAR RP-18 endcapped (2 µm) UHPLC column (Table 1). The performance was compared against the quality control requirements of the GB 31658.17-2021 standard.

Experimental

Sample Preparation

500 g of fresh pork meat was homogenized and sampled by the quarter method into polyethylene bottles.

Reagent Preparation

  1. Sodium dihydrogen phosphate solution (0.05 mol/L): 7.8 g sodium dihydrogen phosphate dihydrate in 1000 mL of water.
  2. Disodium hydrogen phosphate solution (0.05 mol/L): 17.9 g disodium hydrogen phosphate dodecahydrate in 1000 mL of water.
  3. Phosphate buffer solution: 190 mL of 0.05 mol/L sodium dihydrogen phosphate solution top up to 1000 mL with 0.05 mol/L disodium hydrogen phosphate solution.
  4. Sodium hydroxide solution (1 mol/L): 4 g sodium hydroxide in 100 mL of water.
  5. Sodium hydroxide solution (0.03 mol/L): 3 mL of 1 mol/L sodium hydroxide solution topped up to 100 mL with water.
  6. Mcllvaine-Na2EDTA buffer solution: 12.9 g citric acid, 10.9 g disodium hydrogen phosphate, and 39.2 g disodium ethylenediamine tetraacetate were added to 900 mL water and adjusted to pH 5.0 ± 0.2 with 1 mol/L sodium hydroxide solution before topping up to 1000 mL with water.
  7. Eluent (SPE): To 150 mL methanol, add 150 mL ethyl acetate, 6 mL ammonia, and mix well.
  8. Complex solution (reconstitution mix): To 40 mL water, add 5 mL methanol, 5 mL acetonitrile, and 0.05 mL formic acid and mix well.

Standard Preparation

  • Stock solution I (1 mg/mL): Take 10.0 mg of each compound (sulfonamides, quinolones, and tetracycline) in separate 10 mL volumetric flasks. Tetracyclines and sulfonamides were dissolved with methanol and topped to 10 mL. Quinolones were dissolved in 0.03 mol/L sodium hydroxide solution and topped to 10 mL. All solutions were stored below -18°C and to be used within 6 months.
  • Stock solution II (10 µg/mL): Take 0.1 mL of each Stock solution I in separate 10 mL volumetric flasks and top each to mark with methanol. All solutions were stored below -18°C and to be used within 1 month.
  • Stock solution III (1 µg/mL): Take 1 mL of each Stock solution II in separate 10 mL volumetric flasks and top each to mark with methanol. All solutions were stored below -18°C and to be used within 1 month.
  • Calibration standard solutions 1-5: Take 2, 10, 50, 100, and 500 μL of each Stock solution III in a 1 mL volumetric flask and top to mark with methanol for 5 working standard solutions. The concentrations of the drugs are 2.0, 10.0, 50.0, 100, and 500 μg/L.

Sample Preparation

  1. Extraction: To 1 g of the homogenized pork sample, add 8 mL Mcllvaine-Na2EDTA buffer. Vortex for 1 min followed by sonication for 20 min before centrifugation at 10,000 rpm at -2°C for 5 min. Recover the supernatant and set aside. Repeat extraction twice and pool the supernatant as a combined extract.
  2. Solid phase extraction (SPE): Condition and equilibrate the Supel™ Swift HLB SPE Tubes (200 mg/6 mL) tube with 5 mL methanol and 5 mL water. Load the entire pooled supernatant onto the column. Wash the SPE tube with 5 mL of water and 5 mL of 20% v/v methanol. Elute the analytes with 10 mL eluent solution. Collect the eluate and evaporate under nitrogen at 45°C in a water bath. Add 1 mL of the complex solution to reconstitute the residue by swirling for 1 min. Centrifuge at 14,000 rpm for 5 min before filtering through a 0.22 μm PES syringe filter (e.g., Millex®) for LC-MS/MS analysis.
  3. Spiking experiments: Add 50 and 100 μL each of Stock solutions III to 1.0 g pork samples. The resulting spike concentrations of 50 and 100 µg/kg are used for the determination of precision and recovery, respectively.

LC-MS/MS Analysis

The LC-MS/MS analysis of the external calibration standards and the meat extracts was performed with the conditions described in Table 1.

Table 1.LC-MS conditions used for determination of antibiotics in pork meat

Results & Discussion

According to the requirement of the GB method, a gradient elution and detection by multi reaction monitoring (MRM) was used for the LC-MS analysis (Tables 1 & 2). The separation of a 10 µg/L standard solution is displayed in Figure 1 and shows sufficient resolution for the analysis by LC-MS/MS using TIC-MRM in less than 12 min.

A chromatogram obtained in multi reaction monitoring (MRM) mode for a standard solution with a concentration of 10 µg/L. The x-axis represents acquisition time in minutes, ranging from 1 to 12 minutes, while the y-axis displays intensity in counts per second (CPS), with values expressed in multiples of 105. There are multiple peaks of varying heights and retention times, each uniquely colored to differentiate between individual compounds detected in the standard solution. Peaks are closely spaced, with some overlapping, indicating the presence of multiple analytes within a short retention time range. The tallest peaks occur around 5.5 and 11 minutes, representing the highest intensity responses.

Figure 1.Chromatogram (TIC-MRM traces) of a standards solution at 10 μg/L in methanol.

Table 2.Retention times and used MRM transitions for the analysis of a pork sample spiked with 35 drugs.

Calibration

Tetracyclines, sulfonamides and quinolones were externally calibrated (matrix standard solution), and five standard solutions were prepared using a blank pork sample extract. solutions in the range of 2-500 µg/L. The linearity for the 35 compounds showed R2 values between 0.9947 and 0.9998. As an example, in Table 3 and Figure 2, the calibration data for sulfacetamide is shown.

Table 3.Response for sulfacetamide 2, 10, 50, 100 and 500 μg/L standards in methanol.
A graph displaying calibration curve obtained for sulfacetamide at five concentrations of 2, 10, 50, 100 and 500 μg/L. The x-axis represents sulfacetamide concentration in μg/L, ranging from 0 to 500, while the y-axis represents the response value (area), ranging from 0 to 3,000,000. Blue data points, representing different concentrations, are plotted along the curve. A purple dashed line indicates the linear regression fit. The equation of the line is y=56717x+154537, with an R2 value of 0.9998, indicating a strong linear correlation between sulfacetamide concentration and response. This curve is used to quantify unknown sulfacetamide concentrations based on their response values.

Figure 2.Calibration curve for sulfacetamide 2, 10, 50, 100 and 500 μg/L.

Precision and Recovery (%)

The method’s precision was determined by spiked samples at 50 µg/kg and showed %RSDs in the range of 1.0 and 3.5% (n=7) and by that complying with the <15% criteria of the GB method.

The recovery was investigated using samples spiked at 100 µg/kg and showed average recoveries (n=6) between 76.9% and 102%, meeting the 60-110% requirement of the official method.

In Tables 4 & 5 are example data sets for precision assessments and recovery determination for the 3 compounds shown. 

Table 4.Precision data of spiked pork meat sample (50 μg/kg) for 3 selected compounds
Table 5.Recovery experimental data of spiked pork meat sample (100 μg/kg) for 3 selected compounds

Sensitivity

According to the specifications set in the GB31658.17-2021 standard for the determination of Limit of Detection (LOD) & Limit of Quantification (LOQ), this study used the extract of a blank pork sample, where 3N/X was used to determine LOD, 10N/X was used to determine LOQ (Table 6). The GB method states requirements for sensitivity of LOD ≤ μg/kg and LOQ ≤ μg/kg, hence the here-shown method complies with this criteria.

Table 6.LOD & LOQ for determination of tetracyclines, sulfonamides and quinolones (in order of elution)
A TIC-MRM chromatogram obtained for a pork sample spiked with different tetracyclines, sulfonamides and quinolones at concentrations of 10 µg/kg each. The x-axis represents acquisition time in minutes, ranging from 0 to 15 minutes, while the y-axis represents intensity in counts per second (CPS), scaled up to 3.9 × 105. The chromatogram displays a series of sharp peaks at various retention times, indicating the separation and detection of multiple analytes within the spiked pork sample. The peaks vary in height, with some reaching higher intensities. These peaks correspond to the target compounds—tetracyclines, sulfonamides, and quinolones—demonstrating the method's sensitivity in detecting low concentrations of these substances in a complex matrix like pork.

Figure 3.Chromatogram (TIC-MRM) of pork samples spiked with 10 µg/kg of tetracyclines, sulfonamides and quinolones.

TIC-MRM chromatogram of a blank pork sample. The x-axis represents acquisition time in minutes, while the y-axis represents intensity in counts per second (CPS). The chromatogram features a red trace with a series of broad, uneven peaks that rise and fall over time, creating a complex background signal. This trace lacks sharp, distinct peaks, indicating the absence of significant analytes such as tetracyclines, sulfonamides, or quinolones. The signal primarily reflects background noise and matrix-related interferences inherent in the blank pork sample.

Figure 4.Chromatogram (TIC-MRM) of blank pork sample.

Conclusion

35 antibiotic drug residues from tetracyclines, sulfonamides, and quinolones families were analyzed following the GB 31658.17-2021 method. For the SPE cleanup, a Supel™ Swift HLB tube and for the LC-MS/MS analysis a Purospher® STAR RP-18 endcapped (2 µm) UHPLC column were used. The external calibration for the analytes was linear in the range of 2 to 500 μg/L with a correlation coefficient R2 >0.995. The method’s precision was determined by spiked samples at 50 µg/kg and showed % RSDs in the range of 1.0 and 3.5% (n=7). The recovery was investigated using samples spiked at 100 µg/kg and showed average recoveries (n=6) between 76.9% and 102%. The LODs and LOQs for an actual sample were determined according to the GB method with 3.0 μg/kg and 10.0 μg/kg, respectively.

The results of the here-shown method comply with all the criteria set out in the GB standard regarding precision, recovery, and sensitivity (LOD & LOQ). Therefore, the Supel™ Swift HLB SPE tube and the Purospher® STAR RP-18e UHPLC column can be used for the determination of antibiotic residues in food samples according to the GB 31658.17-2021 standard.

Related Products

Loading

References

1.
Liuzhen Y. 2015. Discussion on the status quo, causes and countermeasures of antibiotic abuse in China [J]. . [Internet]. China Science and Technology Journal Database Medicine. Available from: https://xueshu.baidu.com/usercenter/paper/show?paperid=c574b8b51f6e2b4d854ea867a75da583&site=xueshu_se
2.
Ghimpețeanu OM, Pogurschi EN, Popa DC, Dragomir N, Drăgotoiu T, Mihai OD, Petcu CD. Antibiotic Use in Livestock and Residues in Food—A Public Health Threat: A Review. Foods. 11(10):1430. https://doi.org/10.3390/foods11101430
3.
Bhoir SI, Bhoir IC, Bhagwat A, Sundaresan M. 2001. Determination of sulfadoxine in human blood plasma using packed-column supercritical fluid chromatography. Journal of Chromatography B: Biomedical Sciences and Applications. 757(1):39-47. https://doi.org/10.1016/s0378-4347(01)00061-5
4.
Li Y, Zheng Q, Huang Z. 2004. Determination of the content and the related substances of silver sulfadiazine gel by HPLC. Chinese Journal of Hospital Pharmacy. 24(8):474-474.
5.
GB 31658.17-2021 《Determination of tetracycline, sulfonamides and quinolones residues in animal food by liquid chromatography-tandem mass spectrometry》. [Internet]. Available from: https://www.gbstandards.org/