Produktname
Durapore® 0,45 µm, Opticap®-XL-Kapsel, Chemistry: Hydrophilic Polyvinylidene Fluoride (PVDF)Inlet: 3/4 in. Sanitary FlangeOutlet: 3/4 in. Sanitary FlangeMax Operating Pressure: 80 psi @ 25 °C
Quality Segment
material
PVDF , polypropylene , polypropylene housing, polypropylene support, polypropylene vent cap, silicone seal
reg. compliance
meets FDA Indirect Food Additive requirements cited in 21 CFR 177-182 (all component materials)
sterility
non-sterile
sterilization compatibility
autoclavable compatible
product line
Opticap® XL 5
feature
hydrophilic
manufacturer/tradename
Opticap®
parameter
≤7.5 mL/min air diffusion at 1.5 bar (22 psig) and 23 °C (in water), 1.0 bar max. inlet pressure (15 psi) at 80 °C, 2.8 bar max. inlet pressure (40 psi) at 60 °C, 25 °C max. inlet temp., 3.4 bar max. differential pressure (50 psid) at 25 °C (Forward), 3.4 bar max. differential pressure (50 psid) at 25 °C (Reverse; intermittent), 5.5 bar max. inlet pressure (80 psi) at 23 °C, 80 psig max. inlet pressure
technique(s)
bioburden reduction: suitable
L
8.5 in.
W
4.2 in.
cartridge nominal length
5 in. (12.5 cm)
diam.
10.7 cm (4.2 in.)
filtration area
0.34 m2
inlet connection diam.
3/4 in.
inlet to outlet W
21.1 cm (8.3 in.)
outlet connection diam.
3/4 in.
impurities
≤0.5 EU/mL bacterial endotoxins (LAL test, Aqueous extraction), <0.5 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
gravimetric extractables
≤15 mg/capsule
matrix
Durapore®
pore size
0.45 μm
input
sample type liquid
bubble point
≥1930 mbar (28 psig), air with water at 23 °C
fitting
1/4 in. drain/vent hose barb (with double O-ring Seal), inlet sanitary flange, outlet sanitary flange, (19 mm (3/4 in.) Sanitary Flange Inlet and Outlet)
General description
Packaging
Preparation Note
3 Autoklavenzyklen von 60 min bei 126 °C; nicht inline dampfsterilisierbar
Analysis Note
Other Notes
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer Wetting Instructions for Filter Units with Durapore® Membrane user guide
- Storage Statement: Store in dry location
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.
Legal Information
Disclaimer
1 of 1
Dieser Artikel | |||
|---|---|---|---|
| description Chemistry: Hydrophilic Polyvinylidene Fluoride (PVDF)Inlet: 3/4 in. Sanitary FlangeOutlet: 3/4 in. Sanitary FlangeMax Operating Pressure: 80 psi @ 25 °C | description Chemistry: Hydrophilic Polyvinylidene Fluoride (PVDF)Inlet: 1-1/2 in. Sanitary FlangeOutlet: 1-1/2 in. Sanitary FlangeMax Operating Pressure: 80 psi @ 25 °C | description Chemistry: Hydrophilic Polyvinylidene Fluoride (PVDF)Inlet: 9/16 in. Hose BarbOutlet: 9/16 in. Hose BarbMax Operating Pressure: 80 psi @ 25 °C | description Chemistry: Hydrophilic Polyvinylidene Fluoride (PVDF)Inlet: 3/4 in. Sanitary FlangeOutlet: 3/4 in. Sanitary FlangeMax Operating Pressure: 80 psi @ 25 °C |
| pore size 0.45 μm | pore size 0.45 μm | pore size 0.45 μm | pore size 0.1 μm |
| product line Opticap® XL 5 | product line Opticap® XL 5 | product line Opticap® XL 5 | product line Opticap® XL 5 |
| filtration area 0.34 m2 | filtration area 0.34 m2 | filtration area 0.34 m2 | filtration area 0.34 m2 |
| technique(s) bioburden reduction: suitable | technique(s) bioburden reduction: suitable | technique(s) bioburden reduction: suitable | technique(s) sterile filtration: suitable |
| sterility non-sterile | sterility non-sterile | sterility non-sterile | sterility non-sterile |
| matrix Durapore® | matrix Durapore® | matrix Durapore® | matrix Durapore® |
Still not finding the right product?
Explore all of our products under Durapore® 0,45 µm, Opticap®-XL-Kapsel
Bleiben Sie mit dem Emprove®-Programm GMP-konform
Beschleunigen Sie die Qualifizierung, halten Sie die Vorschriften ein und sparen Sie Zeit mit Emprove® Dossiers.
Hinweis: Sie werden zu EmproveSuite.com weitergeleitet. Ihr Benutzername und Passwort sind von SigmaAldrich.com getrennt und können abweichen.
Registrieren Sie sich kostenlos und erhalten Sie innerhalb weniger Minuten Zugang zu Emprove® Dossiers. Aktualisieren Sie Ihr Abonnement, um erweiterte und angepasste Inhalte in der Emprove® Suite freizuschalten.
Sie finden nicht, was Sie suchen? Besuchen Sie die Emprove® Programm-Hilfeseite.
Besitzen Sie dieses Produkt bereits?
In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.
Können Sie Ihr Produkt oder Ihre Lot-/Chargennummer nicht finden?
So finden Sie die Produktnummer
Produktnummern werden bei der Anzeige auf der Website mit Packungsgrößen/-menge kombiniert (Beispiel: T1503-25G). Bitte stellen Sie sicher NUR die Produktnummer in das Feld Produktnummer eingeben (Beispiel: T1503).
Beispiel:
Weitere Beispiele:
705578-5MG-PW
PL860-CGA/SHF-1EA
MMYOMAG-74K-13
1000309185
eingeben als 1.000309185)
Haben Sie Probleme? Sie können sich jederzeit an uns wenden Technischer Dienst zur Unterstützung