Saltar al contenido
Merck
Todas las fotos(1)

Documentos

G1007512

UHPLC/MS Analysis of Insulin Analogs on BIOshell A160 Peptide C18

application for UHPLC-MS

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

technique(s)

UHPLC-MS: suitable

test parameters

column: BIOshell A160 Peptide C18, 15 cm x 2.1 mm I.D., 2.0 μm particles (67243-U)
mobile phase: [A] 75:25 (0.1% DFA in water: 0.1% DFA in acetonitrile); [B] 50:50 (0.1% DFA in water: 0.1% DFA in acetonitrile)
gradient: 23% B for 7 min; 23% B to 100% B in 1 min; hold at 100% B for 7 min
flow rate: 0.2 mL/min
pressure: 2500 psi (172 bar)
column temp.: 75 °C
detector: ESI(+), TIC 100-3000 m/z
injection: 0.5 μL
sample: Analog Mixture, 20 μg/mL, 6:3:1 methanol:water:acetic acid containing 0.1% rat plasma

suitability

application for UHPLC-MS

application(s)

life science and biopharma

Legal Information

Analito

Descripción

    Lantus
    Novolog
    Humalog
    Apidra
    Levemir

Choose from one of the most recent versions:

Certificados de análisis (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico