G005902
HPLC Analysis of Sugars by HILIC Chromatography using Ascentis® Express OH-5 Column with ELSD Detection
application for HPLC
About This Item
técnicas
HPLC: suitable
parámetros analíticos
column: Ascentis Express OH-5, 15 cm x 4.6 mm, I.D., 2.7 μm particles (53778-U)
mobile phase: [A] 0.005 M ammonium formate, pH 3.15 with concentrated TFA [B] acetonitrile (30:70, A:B)
flow rate: 0.5 mL/min
pressure: 841 psi (58 bar)
column temp.: 35 °C
detector: ELSD (detector other: Nebulizer 60º C, Evaporator 80º C, Gas flow 1.60 SLM (Standard Liters/ Minute), LED 100%, Smoothing 5.0 S, PMT ( Photomultiplier) Gain 1.0)
injection: 10 μL
sample: 0.5 mg/mL each prepared in 50:50 Water:Acetonitrile