85-438
HPLC Analysis of Nonvolatile Pesticides in Water on SUPELCOSIL™ LC-18-DB after SPE using Supelclean™ ENVI-Carb™
application for SPE, application for HPLC
About This Item
Recommended Products
technique(s)
HPLC: suitable
solid phase extraction (SPE): suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: water spiked with pesticides at 10-50 μg/L
SPE tube/cartridge: Supelclean ENVI-Carb, 250 mg/3 mL (57088)
condition: 5 mL methylene chloride:methanol (80:20); 1 mL methanol; 10 mL 2% acetic acid in water (keep bed moist until sample addition)
sample addition: 100 mL sample at 5 mL/min
drying: 1 minute with vacuum
elution: 0.8 mL methanol; 2 x 35 mL methylene chloride:methanol (80:20)
eluate post-treatment: dry eluate to 500 μL under gentle nitrogen stream, reconstitute to 1 mL with methanol
column: SUPELCOSIL LC-18-DB, 25 cm × 4.6 mm I.D., 5 μm particles (58355-U)
mobile phase: [A] water:acetonitrile (90:10), [B] acetonitrile; gradient 80% A for 5 min then to 30% A over 30 min
flow rate: 1.5 mL/min
detector: UV, 220 nm
injection: 20 μL
suitability
application for HPLC
application for SPE
application(s)
environmental
Legal Information
Analyte
SPE tube or plate
analytical column
Choose from one of the most recent versions:
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service