Determination of vanillin in commercial food product by adsorptive stripping voltammetry using a boron-doped diamond electrode.

Food chemistry (2013-07-23)
Yavuz Yardım, Mehmet Gülcan, Zühre Şentürk
RESUMEN

A method for the determination of food additive vanillin was developed by adsorptive stripping voltammetry. Its determination was carried out at the anodically pre-treated boron-doped diamond electrode in aqueous solutions. Using square-wave stripping mode, the compound yielded a well-defined voltammetric response in phosphate buffer, pH 2.5 at +1.14 V (vs. Ag/AgCl) (a pre-concentration step being carried out at open-circuit condition for 60s). A linear calibration graph was obtained in the concentration range of 0.5-15.0 μg mL(-1) (3.3×10(-6)-9.8×10(-5) mol L(-1)) with a detection limit of 0.024 μg mL(-1) (1.6×10(-7) mol L(-1)). As an example, the practical applicability of the proposed method was tested for the determination of this flavouring agent in commercial pudding powder of Keshkule (Turkish milk pudding with almond flour).