Saltar al contenido
MilliporeSigma

HPLC Analysis of Flavonoids in Lime Juice on Ascentis® Express C18

HPLC Analysis of Flavonoids in Lime Juice on Ascentis® Express C18 application for HPLC, application for SPE

Materiales

Analytical column

Referencia del producto
Descripción
Precios

Columna para HPLC Ascentis® Express C18, 2,7 μm

2.7 μm particle size, L × I.D. 5 cm × 4.6 mm

mobile phase component

Referencia del producto
Descripción
Precios

2-Propanol

99.5%, HPLC grade

standard

Referencia del producto
Descripción
Precios

Eriocitrin

analytical standard

Hesperidin

analytical standard

Narirutin

≥98% (HPLC)

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

lime juice hand-squeezed in laboratory, centrifuged and filtered through a 0.45 μm nylon filter disc (Acrodisc)

column

Ascentis Express C18, 5 cm x 4.6 mm I.D., 2.7 μm particles (53826-U)

mobile phase

[A] water/formic acid (99.9:0.1); [B] water/acetonitrile/isopropanol/formic acid (39.9:20:40:0.1)

gradient

10% B for 3 min; 10 to 34% B in 30 min; 34 to 100% B in 1 min; held at 100% B for 6 min; 100 to10% B in 1 min

flow rate

0.8 mL/min

column temp.

ambient

detector

PDA: 190–370 nm; chromatogram extracted at 283 nm

detector

ESI-MS: mass spectral range, 250-700 m/z; interval, 0.5 s; scan speed, 938 amu/s; nebulizing gas (N2) flow, 1.5 L/min; (interface temperature, 350 °C; heat block, 300 °C; desolvation line (DL) temperature, 300 °C; DL voltage, 34 V; probe voltage, +4.5 kV; Qarray voltage, 1.0 V; RF voltage, 90 V; detection gain, 1.05 kV)

injection

2 μL

Descripción

Analysis Note

Flavonoids are polyphenolic compounds ubiquitous in nature and classified, according to their chemical structure, into flavonols, flavones, flavanones, isoflavones, catechins, anthocyanidins and chalcones. They are widely present in fruits, vegetables, leaves, nuts, seeds, bark and beverages (tea, coffee, beer, wine and fruit drinks).

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany