Saltar al contenido
MilliporeSigma

US EPA Method OLM04.2 VOA: GC Analysis of Volatiles on VOCOL® after Purge & Trap using "K" Trap, Fast GC Analysis

US EPA Method OLM04.2 VOA: GC Analysis of Volatiles on VOCOL® after Purge & Trap using "K" Trap, Fast GC Analysis suitable for GC

Materiales

analytical column

Referencia del producto
Descripción
Precios

Columna capilar para GC VOCOL®

L × I.D. 20 m × 0.18 mm, df 1.00 μm

CONDITIONS

sample/matrix

each analyte at 50 ppb in 5 mL water

purge trap

VOCARB 3000 “K”

purge

40 mL/min at 25 °C for 11 min

dry purge

2 min

desorption preheat

205 °C

desorption process

150 mL/min at 210 °C for 2 min

bake

260 °C for 10 min

transfer line/valve temp.

110 °C

column

VOCOL, 20 m × 0.18 mm I.D., 1.0 μm (28463-U)

oven

40 °C (0.8 min), 19 °C/min to 125 °C, 32 °C/min to 220 °C (1 min)

inj. temp.

150 °C

MSD interface

200 °C

scan range

m/z = 35-400

carrier gas

helium, 1.4 mL/min

injection

100:1 split

liner

0.75 mm I.D. SPME

Descripción

Legal Information

VOCOL is a registered trademark of Merck KGaA, Darmstadt, Germany