Skip to Content
MilliporeSigma

US EPA Method 608.1/608.2: GC Analysis of Organochlorine Pesticides on SPB®-608

US EPA Method 608.1/608.2: GC Analysis of Organochlorine Pesticides on SPB®-608 suitable for GC

Materials

analytical column

Product No.
Description
Pricing

SPB®-608 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

SPB®-608 Capillary GC Column

L × I.D. 30 m × 0.53 mm, df 0.50

Equity-1701 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

Equity-1701 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 1.00 μm

Equity-1701 Capillary GC Column

L × I.D. 30 m × 0.53 mm, df 0.50 μm

SLB®-5ms Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

SLB®-5ms Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 1.00 μm

SLB®-5ms Capillary GC Column

L × I.D. 15 m × 0.10 mm, df 0.10 μm

accessory

Product No.
Description
Pricing

Wheaton wide-mouth bottle

capacity 32 oz (1000 mL), amber, white polypropylene cap, PTFE flat liner

guard column

Product No.
Description
Pricing

Non-Polar Fused Silica Tubing

L × I.D. 5.0 m × 0.10 mm

Non-Polar Fused Silica Tubing

L × I.D. 5.0 m × 0.25 mm

Non-Polar Fused Silica Tubing

L × I.D. 5.0 m × 0.53 mm

recommended

Product No.
Description
Pricing

Polyimide Sealing Resin

pkg of 5 g

required but not provided

Product No.
Description
Pricing

"Y" GlasSeal Connector, Borosilicate Glass

pkg of 1 ea

"Y" GlasSeal Connector, Fused Silica

pkg of 1 ea

"Y" GlasSeal Connector, Fused Silica

pkg of 3 ea

CONDITIONS

column

SPB-608, 30 m x 0.25 mm I.D., 0.25 μm (24103-U)

oven

75 °C (2 min), 15 °C/min to 300 °C (5 min)

inj. temp.

250 °C

detector

ECD, 300 °C

carrier gas

helium, 1.2 mL/min. constant

injection

1 μL, splitless (splitter open after 0.75 min)

liner

4 mm I.D., single taper

sample

mixture of 608.1 and 608.2 calibration solutions (40351-U and 40352-U) diluted to 100 ppb in n-hexane