HPLC Analysis of a Monoclonal Antibody on Zenix®-C SEC-150 after Partial Proteolysis with IdeS Protease

HPLC Analysis of a Monoclonal Antibody on Zenix®-C SEC-150 after Partial Proteolysis with IdeS Protease application for HPLC

材料

相關產品

產品號碼
描述
訂價

FabRICATOR®

for cleaving 2 mg IgG, 2000 U/vial

CONDITIONS

column

Zenix-C SEC-150, 30 cm X 4.6 mm I.D., 3 μm particles (Z777083)

mobile phase

0.15 M potassium phosphate monobasic, pH 7.0 with potassium hydroxide (567329-U)

flow rate

0.35 mL/min

pressure

1450 psi (100 bar)

column temp.

ambient

detector

UV, 215 nm

injection

1 μL

sample

digested SO57 mAb, ~0.5 g/L

描述

分析報告

Proteolysis of antibodies can be useful for simplifying further analysis. This application demonstrates separation of the two fragments, F(ab′)2 and scFc from the intact IgG, after partial digestion with IdeS protease.