G006151
UHPLC/MS Analysis of Antineoplastic and Antidepressant Drugs in Plasma on Titan C18 after SPE using HybridSPE®-PLus
application for HPLC, application for SPE
About This Item
Recommended Products
technique(s)
HPLC: suitable
solid phase extraction (SPE): suitable
test parameters
sample/matrix: rat plasma
SPE well plate: HybridSPE-PLus 96-Well Plate, 50 mg/2 mL per well (575659-U)
sample preparation: final analyte concentration 50 ng/mL
sample addition: 400 μL of 1% formic acid acetonitile:spiked plasma (3:1) per well
column: Titan C18, 5 cm x 2.1 mm I.D., 1.9 μm particles (577122-U)
mobile phase: [A] 5 mM ammonium formate; [B] 5 mM ammonium formate in 90:10 (v:v) acetonitrile:water
gradient: 35 to 90% B in 1.5 min, held at 90% B for 0.4 min
flow rate: 0.4 mL/min
pressure: 3582 psi (247 bar)
column temp.: 35 °C
detector: TOF/MS
injection: 1 μL
suitability
application for HPLC
application for SPE