跳转至内容
Merck

HPLC Analysis of 16 Triazine and Phenyl Urea Compounds on Ascentis® Express RP-Amide

HPLC Analysis of 16 Triazine and Phenyl Urea Compounds on Ascentis® Express RP-Amide application for HPLC

材料

分析柱

产品编号
说明
价格

Ascentis® Express RP-Amide 2.7μm HPLC 色谱柱

2.7 μm particle size, L × I.D. 15 cm × 3 mm

標準

产品编号
说明
价格

阿特拉津

PESTANAL®, analytical standard

特丁通-去乙基

PESTANAL®, analytical standard

去异丙基阿特拉津

PESTANAL®, analytical standard

异丙隆

PESTANAL®, analytical standard

利谷隆

PESTANAL®, analytical standard

敌草隆

PESTANAL®, analytical standard

绿麦隆

PESTANAL®, analytical standard

草净津

PESTANAL®, analytical standard

特丁通

PESTANAL®, analytical standard

特丁津-去乙基

PESTANAL®, analytical standard

多菌灵

analytical standard

CONDITIONS

column

Ascentis Express RP-Amide, 15 cm x 3 mm I.D., 2.7 μm particles (53919-U)

mobile phase

[A] water; [B] acetonitrile

gradient

5 to 53% B in 70 min

flow rate

0.4 mL/min

column temp.

35 °C

detector

UV, 240 nm

injection

5 μL

sample

20 mg/L (except terbuthylazin desethyl and terbuthylazine 60 mg/L) in 60:40, water: acetonitrile

说明

分析報告

This application demonstrates the suitability of Ascentis Express RP-Amide for the efficient separation of 16 triazine and phenyl urea compounds. Lower column temperature can improve resolution of the critical pair.

法律資訊

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany