技術
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
測試參數
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 10 g homogenized food sample (green onions, apples, Napa cabbage, button mushrooms)
SPE tube/cartridge: Supelclean ENVI-Carb II/PSA SPE Tube (54067-U)
condition: 5 mL acetone:toluene (65:35)
sample addition: 1 mL sample extract
elution: 10 mL of acetone:toluene (65:35)
eluate post-treatment: Evaporate fraction to 0.5 mL then adjust volume to 1 mL with ethyl acetate for GC-NPD analysis
column: SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven: 110 °C (2 min), 2.5 °C/min to 205 °C, 10 °C/min to 310 °C (10 min)
inj. temp.: 250 °C
detector: NPD, 310 °C
carrier gas: helium, 0.9 mL/min constant flow
injection: 2 μL, splitless
liner: 4 mm I.D. dual tapered
適合性
application for SPE
應用
agriculture
food and beverages
準備報告
分析報告
法律資訊
分析物
我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.