Przejdź do zawartości
Merck

UHPLC Gradient Analysis of Boswellic Acids on Titan C18

UHPLC Gradient Analysis of Boswellic Acids on Titan C18 application for UHPLC

Materiały

składnik fazy ruchomej

Numer produktu
Opis
Cennik

Phosphoric acid

≥85 wt. % in H2O, ≥99.999% trace metals basis

CONDITIONS

column

Titan C18, 10 cm x 2.1 mm I.D., 1.9 μm particles (577124-U)

mobile phase

[A] 0.1% phosphoric acid; [B] acetonitrile

gradient

40% B for 0.5 min; to 100% B in 2 min; held at 100% B for 10 min

flow rate

0.4 mL/min

pressure

4050 psi (279 bar)

column temp.

35 °C

detector

UV, 210 nm

injection

2 μL

sample

6 Boswellic Acids 50μg/mL in 70:30, water:methanol

Opis

Informacje prawne

null