Przejdź do zawartości
Merck

GC Analysis of Phenols in Water on a 5% Phenyl Column after SPME using 85 μm Polyacrylate Fiber

GC Analysis of Phenols in Water on a 5% Phenyl Column after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

Materiały

produkt powiązany

Numer produktu
Opis
Cennik

SPME fiber assembly polyacrylate (PA)

df 85 μm(PA, for use with manual holder, needle size 24 ga

CONDITIONS

sample/matrix

50 ppb phenols in water, 27% NaCl, pH 2

SPME fiber

polyacrylate, 85 μm (57304)

extraction

20 min immersed in water with rapid stirring

desorption process

splitless, 280 °C (closed 3 min)

column

5% Phenyl Polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)

oven

40 °C (4 min), 12 °C/min to 260 °C

detector

MS, Scan Range m/z = 45-465 at 0.6 sec/scan

carrier gas

helium, 40 cm/sec at 40 °C