Przejdź do zawartości
Merck

Capillary gas chromatographic analysis of protein amino acids as their N(O,S)-isobutoxycarbonyl methyl ester derivatives.

Biomedical chromatography : BMC (1995-09-01)
S Matsumura, H Kataoka, M Makita
ABSTRAKT

A simple and rapid method for the determination of the 22 protein amino acids by capillary gas chromatography is described. The amino acids were converted into their N(O,S)-isobutoxycarbonyl (isoBOC) methyl ester derivatives and measured by GC with flame ionization detection using a DB-17 capillary column. Arginine was converted into ornithine by arginase treatment prior to the isoBOC reaction. The N(O,S)-isobutoxycarbonylation of amino acids with isobutyl chloroformate was completed within 15 s by sonication in aqueous alkaline media. All of the derivatives were quantitatively resolved as single symmetrical peaks within 9 min by GC on a single column. The detection limits of amino acids were 0.2-4.0 ng per injection, and the calibration curves were linear in the range 0.2-50 micrograms for each amino acid. This method was successfully applied to protein hydrolysate samples. The analytical results of amino acid compositions of several proteins are presented.

MATERIAŁY
Numer produktu
Marka
Opis produktu

Sigma-Aldrich
Isobutyl chloroformate, 98%