The Instructions For Use for NeoClear provides a procedure for use in preparing paraffin embedded sections. Only the paraffin is heated to 60°C. All other steps are performed at room temperature. There is no available information for using NeoClear at 40°C. The Instructions for Use for NeoClear are found on the website's product detail page.
If using for deparaffinization NeoClear is similar in use to Xylene. The user may however need to slightly extend the times for each NeoClear station.
The Instructions for Use state that NeoClear is more sensitive to water. This means it may be necessary to change out the various stations of NeoClear on your staining, compared to changing out Xylene on the staining row.
Wybierz wielkość
| Do Państwa/SKU | Dostępność | Cena netto |
|---|
Informacje o tej pozycji
vapor pressure
6 hPa ( 20 °C)
Quality Segment
form
liquid
autoignition temp.
240 °C
potency
>5000 mg/kg LD50, oral (Rat), >5000 mg/kg LD50, skin (Rabbit)
expl. lim.
0.6 % (v/v)
IVD
for in vitro diagnostic use
greener alternative product characteristics
Safer Solvents and Auxiliaries
Learn more about the Principles of Green Chemistry.
sustainability
Greener Alternative Product
color
clear
refractive index
n20/D 1.418
kinematic viscosity
1.3 cSt(20 °C)
mp
-20 °C
transition temp
flash point 43 °C
density
0.78-0.795 g/cm3 at 20 °C
application(s)
diagnostic assay manufacturing
hematology
histology
greener alternative category
storage temp.
15-25°C
General description
Legal Information
1 of 1
Ta pozycja | |||
|---|---|---|---|
| Quality Level 400 | Quality Level - | Quality Level 400 | Quality Level 100 |
| application(s) diagnostic assay manufacturing | application(s) hematology | application(s) diagnostic assay manufacturing | application(s) - |
| form liquid | form solution | form - | form liquid |
| mp -20 °C | mp - | mp - | mp -26--23 °C (lit.) |
| expl. lim. 0.6 % (v/v) | expl. lim. - | expl. lim. - | expl. lim. 7 % |
| vapor pressure 6 hPa ( 20 °C) | vapor pressure - | vapor pressure - | vapor pressure <0.1 atm ( 21.1 °C), 16 mmHg ( 37.7 °C), 7 mmHg ( 20 °C) |
signalword
Danger
hcodes
Hazard Classifications
Aquatic Chronic 4 - Asp. Tox. 1 - Flam. Liq. 3
supp_hazards
Klasa składowania
3 - Flammable liquids
wgk
WGK 1
flash_point_f
110.3 °F
flash_point_c
43.5 °C
Certyfikaty analizy (CoA)
Poszukaj Certyfikaty analizy (CoA), wpisując numer partii/serii produktów. Numery serii i partii można znaleźć na etykiecie produktu po słowach „seria” lub „partia”.
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
-
Are there any experiences with adipose/fat tissue and NeoClear? How long should be the incubation times on 40°C?
1 answer-
Helpful?
-
-
Could this melt the polycarbonate membranes just like in the case of xylene?
1 answer-
The product's chemical composition is "hydrocarbons, C10-C12, Isoalkanes at 100% or less. It is not known whether Neo-Clear is compatible with Polycarbonate.
Helpful?
-
-
Is it possible to use NeoClear with paraffin tissue that was initially processed using Xylene?
1 answer-
After embedding the slides in paraffin, Neoclear or Xylene can be used to clear the slides and remove the paraffin before staining. If NeoClear is chosen, NeoMount should serve as the permanent mounting media. It is generally not recommended to use NeoClear and a xylene-based permanent mounting media, as the standard practice in histology is to use a clearing agent that matches the permanent mounting media. If an aqueous mounting media is utilized, the slides are stained, washed in water, and then coverslipped with the aqueous mounting media. In the case of aqueous mounting media, the slides do not undergo NeoClear or Xylene treatment.
Helpful?
-
-
Is the solvent Neo Clear 1.09843.5000 compatible with a xylene-based mounting medium, or there is any recommendation of any other mounting medium?
1 answer-
Neo Clear is best used with Neo Mount, while Xylene should be avoided.
Helpful?
-