Saltar al contenido
Merck

HPLC Analysis of Methionine Enantiomers on Astec® CHIROBIOTIC® T

HPLC Analysis of Methionine Enantiomers on Astec® CHIROBIOTIC® T application for HPLC

Materiales

Analytical column

Referencia del producto
Descripción
Precios

Columna quiral para HPLC Astec® CHIROBIOTIC® T

5 μm particle size, L × I.D. 25 cm × 4.6 mm

Guard cartridge

Referencia del producto
Descripción
Precios

Astec® Chirobiotic® T Chiral HPLC Guard Column

Cartridge, 5 μm particle size, L × I.D. 2 cm × 4 mm

Standard

Referencia del producto
Descripción
Precios

D-Methionine

≥98% (HPLC)

L-Metionina

from non-animal source, meets EP, JP, USP testing specifications, suitable for cell culture, 99.0-101.0%

L-Metionina

reagent grade, ≥98% (HPLC)

DL-Metionina

BioReagent, suitable for cell culture, suitable for insect cell culture, ≥99%

L-Metionina

L-Metionina

BioUltra, ≥99.5% (NT)

DL-Metionina

≥99.0% (NT)

DL-Metionina

99%, FCC, FG

L-Metionina

certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland

CONDITIONS

column

CHIROBIOTIC T, 25 cm x 4.6 mm I.D., 5 μm particles (12024AST)

mobile phase

(A) water; (B) methanol; (C) formic acid; (30:70:0.02, A:B:C)

flow rate

1 mL/min

column temp.

25 °C

detector

UV, 205 nm

injection

10 μL

sample

300 μg /mL in 30:70 water:methanol