Saltar al contenido
Merck

Volatile compounds by US EPA 524.2

Volatile compounds by US EPA 524.2 suitable for GC, application for SPME

Materiales

analytical column

Referencia del producto
Descripción
Precios

Columna capilar para GC VOCOL®

L × I.D. 30 m × 0.25 mm, df 1.50 μm

SPME fiber

Referencia del producto
Descripción
Precios

Fibra de carboxeno/polidimetilsiloxano (DVB/CAR/PDMS) de SPME

df 75 μm(CAR/PDMS, for use with manual holder, needle size 24 ga

standard

Referencia del producto
Descripción
Precios

EPA 524.2 VOC Mix

200 μg/mL each component in methanol, analytical standard

CONDITIONS

sample/matrix

1 ppb each analyte in 3 mL water (25% NaCl added) in 4 mL vial

SPME fiber

Carboxen/PDMS, 75 μm (57318)

extraction

headspace, 40°C (20 min), with stirring

desorption process

5 min., 310 °C

column

VOCOL®, 30 m × 0.25 mm I.D., 1.5 μm (24205-U)

oven

40°C (2 min) to 210°C at 8°C/min

detector

ion trap, MS, m/z = 45-260 (0.6sec/scan)

carrier gas

helium, 35 cm/sec

injection

splitless/split (closed 2 min), 0.75 mm I.D. liner

Descripción

Legal Information

VOCOL is a registered trademark of Merck KGaA, Darmstadt, Germany