Skip to Content
Merck

HPLC Analysis of Protein Standards on Zenix® SEC-300 by Size Exclusion As Affected by Potassium Chloride

HPLC Analysis of Protein Standards on Zenix® SEC-300 by Size Exclusion As Affected by Potassium Chloride application for HPLC

Materials

analytical column

Product No.
Description
Pricing

Zenix® SEC-300 Gel Filtration Column, 3 μm

L × I.D. 30 cm × 4.6 mm, 3 μm particle size, stainless steel

accessory

Product No.
Description
Pricing

EXP® Pre-Column Filter

holder with EXP titanium hybrid ferrule, pack of 1 sets

EXP® Pre-Column Filter Cartridges

0.5μm, pkg of 5 ea

mobile phase component

Product No.
Description
Pricing

Potassium chloride

BioXtra, ≥99.0%

standard

Product No.
Description
Pricing

Myoglobin from equine heart

≥90% (SDS-PAGE), essentially salt-free, lyophilized powder

Uracil

≥99.0%

Thyroglobulin from bovine thyroid

(For Gel Filtration Chromatography)

IgG from rabbit serum

reagent grade, ≥95% (SDS-PAGE), essentially salt-free, lyophilized powder

Bovine Serum Albumin

lyophilized powder, ≥96% (agarose gel electrophoresis)

CONDITIONS

column

Zenix SEC-300, 30 cm x 4.6 mm I.D., 3 μm particles (Z777028)

mobile phase

[A], 0.15 M potassium phosphate, monobasic, pH 7.0 with potassium hydroxide; [B] 0.15 M potassium phosphate, monobasic, 0.2 M potassium chloride, pH 7.0 with potassium hydroxide; (mobile phase is either 100% A or 100% B)

flow rate

0.3 mL/min

pressure

1000 psi (69 bar)

column temp.

25 °C

detector

UV, 215 nm

injection

2.5 μL

sample

Waters BEH200 SEC protein standard

Description

Analysis Note

In size exclusion chromatography, addition of halide salts (typically sodium or potassium chloride) in modest concentration, to an aqueous buffer, can yield improved chromatographic performance with protein analytes. In this case, as regards chromatographic performanc, little impact is noted except in the case of IgG, where peak shape is improved.

Legal Information

Zenix is a registered trademark of Sepax Technologies