Saltar al contenido
Merck

HPLC Analysis of mAb Aggregates on TSKgel® UltraSW Aggregate: Effect of Mobile Phase Additives on Resolution

HPLC Analysis of mAb Aggregates on TSKgel® UltraSW Aggregate: Effect of Mobile Phase Additives on Resolution application for HPLC

Materiales

analytical column

Referencia del producto
Descripción
Precios

Columna de HPLC de exclusión por tamaño (tipo SW) TSKgel®

phase diol, L × I.D. 30 cm × 7.8 mm, 3 μm particle size

guard column

Referencia del producto
Descripción
Precios

TSKgel® UltraSW Aggregate Guard Column

phase diol, L × I.D. 4 cm × 6 mm, 3 μm particle size

CONDITIONS

column

TSKgel UltraSW Aggregate 30 cm x 7.8 mm I.D., 3 μm particles (822856)

mobile phase

0.1 mol/L sodium phosphate buffer, pH 6.7, with an addition of either 0.2 mol/L arginine (A), or 0.2 mol/L proline (B)

flow rate

1.0 mL/min

column temp.

ambient

detector

UV, 280 nm

injection

20 μL

sample

mAB, aggregated at 75 °C for 5 min

Descripción

Analysis Note

While particle size, packing quality and column length are all important considerations in optimizing mAb aggregate analysis, mobile phase additives can also contribute to improving your separation. The addition of amino acids such as lysine and arginine to the mobile phase, increase aggregate recovery, while glutamine, proline and arginine improve monomer aggregate resolution, as well as monomer fragment resolution. Although the increases in resolution are not drastic, they confirm that increased resolution due to the use of an advanced mobile phase is possible and that mobile phase testing can contribute to a more reliable and robust aggregate analysis.

Legal Information

TSKgel is a registered trademark of Tosoh Corporation