Modification of the selected method for the determination of serum iron. Substitution of bathophenanthroline by ferene S.

Annales de biologie clinique (1992-01-01)
C Charlier, G Plomteux, M Vernet, P Gendre, M C Revenant, C Guillemin
RESUMEN

The previously selected method (1977) for the measurement of iron in human serum has been modified by the working group on Iron and Iron-Transport Proteins (Sociéte Française de Biologie Clinique). Ferene S is used as the chromogen, it is more sensitive and cheaper than bathophenanthroline. The sample size has been reduced from 2 to 0.5 ml. No interference could be shown with high concentrations of bilirubin, copper or hemoglobin.

MATERIALES
Referencia del producto
Marca
Descripción del producto

Supelco
3-(2-Pyridyl)-5,6-di(2-furyl)-1,2,4-triazine-5′,5′′-disulfonic acid disodium salt, for spectrophotometric det. of Fe, ≥99.0%
Sigma-Aldrich
3-(2-Pyridyl)-5,6-di(2-furyl)-1,2,4-triazine-5′,5′′-disulfonic acid disodium salt, Suitable for determination of Fe