Saltar al contenido
Merck

Attomole protein analysis by CIEF with LIF detection.

Electrophoresis (2009-02-11)
Lauren M Ramsay, Jane A Dickerson, Norman J Dovichi
RESUMEN

We have coupled CIEF with an LIF detector that is based on a post-column sheath flow cuvette. We employed Chromeo P503 as a fluorogenic reagent to label proteins before analysis. This reagent reacts with the epsilon-amine of lysine residues, preserving the cationic nature of the residue; labeled proteins generate extremely sharp peaks in CIEF. A set of four standard proteins generated a linear relationship between migration time and pI. A protein homogenate prepared from a Barrett's esophagus cell line resolved over 100 components in a 40 min separation. Detection limits for Chromeo P503-labeled beta-lactoglobulin were 5 amol injected into the capillary. Fluorescent impurities present in the ampholytes generated a large background signal that degraded the detection limit by four orders of magnitude compared with other forms of capillary electrophoresis with this detector.

MATERIALES
Referencia del producto
Marca
Descripción del producto

Sigma-Aldrich
Chromeo P503