APP501
HPLC Analysis of Cinacalcet Hydrochloride Impurities with a Purospher® STAR Phenyl Column
About This Item
technique(s)
HPLC: suitable
test parameters
column: Purospher® STAR Phenyl (5 µm) LiChroCART® 150 x 4.6 mm I.D. (1.51922)
mobile phase: [A] 10 mmol/L Sodium perchlorate solution, adjusted to pH 2.5 with perchloric acid; [B] Acetonitrile
gradient: 0-0.5 min 2.5% B, in 4.5 min to 95%B, 2 min 95%B, 3 min at 2.5%B
flow rate: 1.0 mL/min
pressure: 60 bar (870 psi)
column temp.: 60°C (Sampler: 10 °C)
detector: UV, 215 nm (analytical flow cell; 10 µL)
injection: 5 µL
sample/matrix: Impurity identification solution with cinacalcet impurities A, B, C at 1.5 µg/mL each and cinacalcet at 5 µg/mL in solvnt mixture (10 mmol/L sodium perchlorate solution, adjusted to pH 2.5 with perchloric acid, and acetonitrile in a ratio of 50:50 (%v/v))
application(s)
pharmaceutical