US EPA Method 8015 (modified): GC Analysis of Diesel Range Organics (DRO) on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm), Fast GC Analysis
Materials
analytical column
Product No.
Description
Pricing
accessory
Product No.
Description
Pricing
Wheaton wide-mouth bottle
amber soda-lime glass bottle, capacity 1000 mL, white polypropylene cap, (PTFE faced LDPE liner), case of 12 eaWheaton straight-mouth bottles with cap, clear
clear soda-lime glass bottle, capacity 250 mL, white polypropylene cap, (PTFE faced LDPE liner), case of 12 eaguard column
recommended
Product No.
Description
Pricing
required but not provided
standard
Product No.
Description
Pricing
Underground Storage Tank (UST) Modified DRO
certified reference material, TraceCERT®, 1000 μg/mL each component in hexane, ampule of 1 mLCONDITIONS
column
SLB-5ms, 20 m x 0.18 mm I.D., 0.18 μm (28564-U)
oven
50 °C (2 min), 20 °C/min to 325 °C (1 min)
inj. temp.
250 °C
detector
FID, 350 °C
carrier gas
helium, 0.9 mL/min, constant
injection
1 μL, splitless (splitter open at 0.5 min)
liner
2 mm I.D., FocusLiner™ inlet liner
sample
each analyte at 50 μg/mL in methylene chloride
Description
Legal Information
FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany