Skip to Content
Merck

The biomimetic apatite-cefalotin coatings on modified titanium.

Dental materials journal (2012-01-27)
Min-Kyung Kang, Sang-Bae Lee, Seung-Kyun Moon, Kwang-Mahn Kim, Kyoung-Nam Kim
ABSTRACT

Dental implant failure often occurs due to oral bacterial infection. The aim of this study was to demonstrate that antibiotic efficacy could be enhanced with modified titanium. First, the titanium was modified by anodization and heat-treatment. Then, a biomimetic coating process was completed in two steps. Surface characterization was performed with scanning electron microscopy, energy dispersive spectroscopy, and X-ray diffraction. Release of antibiotic was evaluated by UV/VIS spectrometry, and the antibacterial effect was evaluated on Streptococcus mutans. After the second coating step, we observed a thick homogeneous apatite layer that contained the antibiotic, cefalotin. The titanium formed a rutile phase after the heat treatment, and a carbonated apatite phase appeared after biomimetic coating. We found that the modified titanium increased the loading of cefalotin onto the hydroxyapatite coated surface. The results suggested that modified titanium coated with a cefalotin using biomimetic coating method might be useful for preventing local post-surgical implant infections.

MATERIALS
Product Number
Brand
Product Description

Sigma-Aldrich
Cephalothin sodium salt