콘텐츠로 건너뛰기
Merck
모든 사진(1)

문서

G005882

LC/MS/MS Analysis of Testosterone and Metabolites on Ascentis® Express C18

application for HPLC

로그인조직 및 계약 가격 보기


About This Item

기술

HPLC: suitable

검사 파라미터

column: Ascentis Express C18, 5 cm x 2.1 mm I.D., 2.7 μm particles (53822-U)
mobile phase: [A] 5 mM ammonium formate water pH 4.0 formic acid; [B] 0.1% formic acid acetonitrile
gradient: 10 to 70% B in 5 min; held at 70% B for 3 min
flow rate: 0.6 mL/min
pressure: 3698 psi (255 bar)
column temp.: 35 °C
detector: MS, ESI(+), TIC, m/z 100-1000
injection: 5 μL
sample: 1000 ng/mL in 40:60 (water:methanol)

적합성

application for HPLC

응용 분야

clinical

분석 메모

An Ascentis Expres F5 column provided rapid, baseline resolution of testosterone and metabolites. The highest quality LC-MS solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.

법적 정보

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

분석물

설명

    Testosterone solution, 1.0 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®
    17α-Methyltestosterone solution, 1.0 mg/mL in 1,2-dimethoxyethane, ampule of 1 mL, certified reference material, Cerilliant®
    5α-Dihydrotestosterone (DHT) solution, 1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®
    6β-Hydroxytestosterone solution, 100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Choose from one of the most recent versions:

시험 성적서(COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.