Determination of Chlorogenic acid and Baicalin in Compounded Lonucerae Japonicae Flos acc. Chinese Pharmacopeia using a Discovery® HS C18 Column
Dean Duan
Merck, Application Scientist, China application lab, Shanghai, China
Section Overview
Introduction
Compounded Lonucerae Japonicae Flos is a kind of Traditional Chinese medicine (TCM). It has the effects of heat-clearing and detoxicating, cooling blood and detumescence. In this application, one Chinese pharmacopeia 2020 monograph method was validated using a Discovery® HS C18 HPLC column.
![Chemical structures of chlorogenic acid and baicalin 2-D chemical structures (bond line structure) of chlorogenic acid and baicalin used to measure analytes in compounded Lonucerae Japonicae Flos using a Discovery® HS C18](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/chemical-structures-of-chlorogenic-acid-and-baicalin/chemical-structures-of-chlorogenic-acid-and-baicalin.jpg)
Figure 1. Chemical structures of chlorogenic acid and baicalin.
![HPLC profile of chlorogenic acid and baicalin standards Analysis of chlorogenic acid and baicalin standards used to determine the analytes in compounded Lonucerae Japonicae Flos using a Discovery® HS C18 HPLC column](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/chromatogram-of-chlorogenic-acid-and-baicalin-standard-solution/chromatogram-of-chlorogenic-acid-and-baicalin-standard-solution.jpg)
Figure 2. Chromatogram of chlorogenic acid and baicalin standard solution.
![HPLC profile of compounded Lonucerae Japonicae Flos Analysis of compounded Lonucerae Japonicae Flos using Discovery® HS C18 HPLC column](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/chromatogram-of-compounded-lonucerae-japonicae-flos/chromatogram-of-compounded-lonucerae-japonicae-flos.jpg)
Figure 3. Chromatogram of compounded Lonucerae Japonicae Flos.
![HPLC of solvent blank Separation of solvent blank on Discovery® HS C18 HPLC column](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/chlorogenic-acid-and-baicalin-/hplc-of-solvent-blank/hplc-of-solvent-blank.jpg)
Figure 4. Chromatogram of solvent blank.
Chromatograms at 326 nm for baicalin
![HPLC profile of chlorogenic acid and baicalin standard solution Chromatographic separation of chlorogenic acid and baicalin standards in compounded Lonucerae Japonicae Flos using Discovery® HS C18 HPLC column](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/chlorogenic-acid-and-baicalin-/hplc-profile-of-chlorogenic-acid-and-baicalin-standard-solution/hplc-profile-of-chlorogenic-acid-and-baicalin-standard-solution.jpg)
Figure 5. Chromatogram of chlorogenic acid and baicalin standard solution.
![HPLC profile of compounded Lonucerae Japonicae Flos Chromatogram of compounded Lonucerae Japonicae Flos to analyze chlorogenic acid and baicalin analytes using Discovery® HS C18 HPLC column](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/chlorogenic-acid-and-baicalin-/hplc-profile-of-compounded-lonucerae-japonicae-flos/hplc-profile-of-compounded-lonucerae-japonicae-flos.jpg)
Figure 6. Chromatogram of compounded Lonucerae Japonicae Flos.
![HPLC of solvent blank Separation of solvent blank on Discovery® HS C18 HPLC column](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/chlorogenic-acid-and-baicalin-/hplc-of-solvent-blank/hplc-of-solvent-blank.jpg)
Figure 7. Chromatogram of solvent blank.
Specificity and repeatability – chlorogenic acid and baicalin
Calibration
Calibration data – chlorogenic acid
![Calibration curve of chlorogenic acid Calibration curve of chlorogenic acid used to determine the analyte in compounded Lonucerae Japonicae Flos with Discovery® HS C18 HPLC column](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/chlorogenic-acid-and-baicalin-/calibration-graph-of-chlorogenic-acid/calibration-graph-of-chlorogenic-acid.jpg)
Figure 8. Calibration curve of chlorogenic acid.
Calibration data – baicalin
![Calibration graph of baicalin Calibration curve of baicalin used to determine the analyte in compounded Lonucerae Japonicae Flos with Discovery® HS C18 HPLC column](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/chlorogenic-acid-and-baicalin-/calibration-graph-of-baicalin/calibration-graph-of-baicalin.jpg)
Figure 9. Calibration curve of baicalin
Conclusion
The Chinese pharmacopeia 2020 monograph method was reproduced using a Discovery® HS C18 HPLC column. For the shown method, the limit of detection (LOD) and the limit of quantitation (LOQ) were found to be 0.07 µg/mL and 0.21 µg/mL, respectively, for chlorogenicacid, while the limit of detection (LOD) and the limit of quantitation (LOQ) were 0.27 µg/mL and 0.82 µg/mL, respectively, for baicalin. The method can be used for determination of chlorogenicacid and baicalin in compounded Lonucerae Japonicae Flos according to the Chinese pharmacopoeia 2020.