コンテンツへスキップ
Merck

P2B005V20

Millipore

Pellicon® 2 MaxiウルトラフィルトレーションBiomax® モジュール

filtration area 2 m2, NMWCO 5 kDa

別名:

Pellicon® 2 Maxi Ultrafiltration Module Biomax-5 V 2 m2

ログイン組織・契約価格を表示する


About This Item

UNSPSCコード:
23151806

物質

polyethersulfone
polypropylene screen
polyurethane adhesive

品質水準

メーカー/製品名

Pellicon®

パラメーター

0.33 bar max. transmembrane pressure (5 psig) (reverse)
5-35 L/min recirculation rate at 0.4 bar (6 psi)
7 bar max. transmembrane pressure (100 psig) at 30 °C (forward)

テクニック

analytical sample preparation: suitable
protein purification: suitable

H

9.0 cm (3.5 in.)

L

21 cm (8.3 in.)

W

17.8 cm (7.0 in.)

ろ過面積

2 m2

membrane nominal area

2 m2

重量分析抽出物

≤1500 mg/m2

マトリックス

Biomax®

ポアサイズ

5 kDa NMWCO

pH

1-14

類似した製品をお探しですか? 訪問 製品比較ガイド

詳細

The Pellicon® 2 Ultrafiltration BIOMAX 5 Maxi Cassette sets the standard in cassette-style tangential flow filtration devices. P2B005V20 was designed with process scale-up and applications development in mind.

Millipore′s BIOMAX 5 polyethersulfone membrane has void-free structures that guard against leakage of solutes through microdefects normally associated with voids beneath the thin skins of conventional UF membranes. These void-free membranes are more permeable, resulting in high-flux with equivalent or superior product retention. These void-free membranes provide the advantages of fast, high yield processing and smaller systems.

The Pellicon® 2 Maxi Ultrafiltration BIOMAX Module feature a filtration area of 2 m2 with a pore size of 5 kDa NMWCO.

アプリケーション

Concentration, desalting or buffer exchange of:
  • Protein solutions
  • Polysaccharide solutions
  • Virus suspensions


Harvest, washing or clarification of:
  • Cell cultures and lysates
  • Colloidal suspensions
  • Viral cultures

特徴および利点

  • Type V screen (open channel)
  • Leading-edge void- free membranes to match virtually any separation challenge
  • Short flow path for higher flux and higher resolution separation capability
  • Choice of flow channel configuration providing process optimization capability
  • Predictable, fast, scale-up
  • True linear scalability from laboratory size modules to industrial assemblies for processing thousands of liters
  • Brings higher resolution, improved yields and superior back-pressure resistance
  • Ultracel membranes are composite membranes cast on a microporous substrate for defect-free membranes with superior

その他情報

これらの製品はFDAの定めたGMP準拠の製造工場で製造されています。
Prefiltration Rquirements: 100 µm

アナリシスノート

Gravimetric Extractables: after a water flush

法的情報

BIOMAX is a registered trademark of Merck KGaA, Darmstadt, Germany
Pellicon is a registered trademark of Merck KGaA, Darmstadt, Germany

保管分類コード

10 - Combustible liquids

WGK

WGK 2


適用法令

試験研究用途を考慮した関連法令を主に挙げております。化学物質以外については、一部の情報のみ提供しています。 製品を安全かつ合法的に使用することは、使用者の義務です。最新情報により修正される場合があります。WEBの反映には時間を要することがあるため、適宜SDSをご参照ください。

労働安全衛生法名称等を通知すべき危険物及び有害物

名称等を通知すべき危険物及び有害物

Jan Code

P2B005V20:


試験成績書(COA)

製品のロット番号・バッチ番号を入力して、試験成績書(COA) を検索できます。ロット番号・バッチ番号は、製品ラベルに「Lot」または「Batch」に続いて記載されています。

以前この製品を購入いただいたことがある場合

文書ライブラリで、最近購入した製品の文書を検索できます。

文書ライブラリにアクセスする

ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.

製品に関するお問い合わせはこちら(テクニカルサービス)