UHPLC Analysis of Bisphenol A and Analogous Compounds in Infant Formula on Titan C18 with Fluorescence Detection after SPE using SupelMIP® Bisphenol A

UHPLC Analysis of Bisphenol A and Analogous Compounds in Infant Formula on Titan C18 with Fluorescence Detection after SPE using SupelMIP® Bisphenol A application for UHPLC, application for SPE

Materiali

Accessorio

N° Catalogo
Descrizione
Determinazione del prezzo

Screw cap (open-top), 9 mm thread, with liner

clear cap, PTFE/silicone septum, pkg of 100 ea

ABC Vial, 2 mL (vial only), 9 mm thread

clear glass vial (with marking spot), pkg of 100 ea

Colonnina o piastra per SPE

N° Catalogo
Descrizione
Determinazione del prezzo

SupelMIP® SPE - Bisphenol A

bed weight 100 mg, volume 6 mL, pk of 50

Preparazione del campione

N° Catalogo
Descrizione
Determinazione del prezzo

Supel QuE

Empty Centrifuge Tube with Lid, centrifuge tube volume 50 mL, pk of 50, suitable for EN 15662:2008 per BS

Standard

N° Catalogo
Descrizione
Determinazione del prezzo

Bisphenol A diglycidyl ether

Bis(4-hydroxyphenyl)methane

98%

Bisphenol A

≥99%

Bisphenol C

analytical standard

Bisphenol G

analytical standard

CONDITIONS

sample/matrix

liquid infant formula

SPE well plate

SupelMIP Bisphenol A, 100 mg/3 mL (52775-U)

condition

3 mL 2% formic acid in methanol, 3 mL acetonitrile, 3 mL water

sample addition

as prepared

washing

9 mL of water, 6 mL of acetonitrile:water (40:60); vacuum was pulled for 3 min after the last wash

eluate post-treatment

Sample was evaporated to dryness at 45 °C, reconstuituted in 0.5 mL of acetonitrile:water (40:60). All samples were filtered through Durapore membrane filter units prior to analysis.

column

Titan C18, 10 cm x 2.1 mm I.D., 1.9 μm particles (577124-U)

mobile phase

[A] water; [B] acetonitrile

gradient

40 to 80% B in 2.5 min; held for 4.5 min; to 100%B in 0.1 min; held for 2.4 min; to 40% B in 0.1 min; held for 2.9 min

flow rate

0.25 mL/min for 7 min; to 0.40 mL/min in 0.1 min; held for 2.4 min; to 0.25 mL/min in 0.1 min.; held for 2.9 min

pressure

3916 psi (270 bar)

column temp.

35 °C

detector

fluorescence, ex: 230 nm, em: 315 nm

injection

5 μL

elution

3 mL of methanol:acetonitrile (75:25) (A017)

Descrizione

Risultati analitici

BPA can leach into foods from the contact between food and the packaging. Structurally analogous compounds could also be used in place of BPA during package manufacturing as BPA comes under more stringent regulations. The presented method was developed for analysis of this class of compounds and applied to infant formula analysis. The use of molecularly imprinted polymers during cleanup produced samples that can be analysed by fluoresent detection.

Altre note

Sample Prep: 5 mL of liquid infant formula was spiked with 10 ppb of analytes, mixed with 5.5 mL of acetonitrile, centrifuged at 4500 rpm for 10 minutes, top 3 mL were diluted with 3 mL water to load into SPE