Skip to Content
Merck

GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) in Grilled Hamburger on SPB®-608 (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel QuE Z-Sep, Fast GC Analysis

GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) in Grilled Hamburger on SPB®-608 (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel™ QuE Z-Sep, Fast GC Analysis suitable for GC, application for SPE

Materials

SPE tube or plate

Product No.
Description
Pricing

Supel QuE

Non-Buffered Tube 1, pk of 50

related product

Product No.
Description
Pricing

Supel QuE

Empty Centrifuge Tube with Lid, centrifuge tube volume 50 mL, pk of 50, suitable for EN 15662:2008 per BS

CONDITIONS

sample preparation

Dispersive (QuEChERS)

sample/matrix

grilled hamburger spiked with PAHs at 100 ng/g

extraction tube

Sodium chloride/magnesium sulfate (55294-U)

extraction process

add 25 mL acetonitrile, contents of unbuffered salt tube (55294-U); shake for 5 minutes; centrifuge for 5 minutes at 3200 rpm

clean-up tube

Supel QuE Z-Sep tube (55403-U)

clean-up process

transfer 3 mL of acetonitrile layer into a Supel QuE Z-Sep cleanup tube (55403-U); shake 1 minute; centrifuge for 3 minutes at 3400 rpm; remove 1 mL supernatant for GC-MS analysis

column

SPB-608, 20 m x 0.18 mm I.D. x 0.18 μm

oven

60 °C (1 min), 25 °C/min to 275 °C, 10 °C/min to 300 °C (13 min)

inj. temp.

265 °C

detector

MS-SIM

MSD interface

300 °C

carrier gas

helium, 1.5 mL/min constant flow

injection

1 μL splitless (1 min.)

liner

4 mm I.D. FocusLiner with taper

sample

grilled hamburger extract

Description

Analysis Note

Sample pretreatment: Place 5 g ground hamburger sample into 50 mL centrifuge tube (55248-U). Spike with 100 ng/g target PAHs and internal standards. Let sit 10 min. Add 5 mL water and homogenize.
This application demonstrates the use of dispersive SPE or QuEChERS for the extraction and cleanup of grilled hamburger for GC-MS analysis of polynuclear aromatic hydrocarbons (PAHs). Z-Sep sorbent was used in the cleanup step and found to reduce background and yield recoveries >80% for a majority of the PAHs tested.

Other Notes

Legal Information

FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SPB is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC