- Automated carboxylesterase assay for the evaluation of ionic liquids' human toxicity.
Automated carboxylesterase assay for the evaluation of ionic liquids' human toxicity.
Aiming the prediction of ionic liquids' (ILs) human toxicity, an automated carboxylesterase activity assay was developed. The method was implemented on a sequential injection analysis (SIA) system and relied on the hydrolysis of 4-methylumbelliferyl acetate by the enzyme, to produce the fluorescent compound 4-methylumbelliferone (λ(exc)=365 nm; λ(em)=460 nm). The adopted approach was based on activity/inhibition assays in which the reduction of enzyme activity was denoted by a decrease of fluorescence. The activity assays were performed in aqueous media and in the presence of increasing concentrations of seven commercially available ILs and sodium bis(p-nitrophenyl) phosphate, a strong inhibitor of carboxylesterase. The inhibitory effects were quantified by means of the calculation of half maximal inhibitory effective concentration (EC(50)) values that were between 38 and 134 μmol L(-1). This allowed us to perform some considerations about the toxicity of the compounds. The less and more toxic ILs found in this study were emim [Ms] and tbph [Ms], respectively. The developed SIA methodology is robust and exhibits good repeatability in all the tested conditions (rsd<5.0%, n=10). At the same time a reduction of costs was attained by reduced consumption of reagents and minimal effluent production (2 mL per cycle).