Saltar al contenido
Merck

Using the column wall itself as resistive heater for fast temperature gradients in liquid chromatography.

Journal of chromatography. A (2015-10-20)
Ruben De Pauw, Matthias Pursch, Gert Desmet
RESUMEN

A new system is proposed for applying fast temperature gradients in liquid chromatography. It consists of a 0.7 mm × 150 mm fused-silica column coated with a 50 μm Nickel-layer, which is connecting with a power source and a temperature control system to perform fast and reproducible temperature gradients using the column wall itself as a resistive heater. Applying a current of 4A and passive cooling results in a maximal heating and cooling rate of, respectively, 71 and -21 °C/min. Multi-segment temperature gradients were superimposed on mobile phase gradients to enhance the selectivity for three sets of mixtures (pharmaceutical compounds, a highly complex mixture and an insecticide sample). This resulted in a higher peak count or better selectivities for the various mixtures.

MATERIALES
Referencia del producto
Marca
Descripción del producto

Sigma-Aldrich
Acetofenona, ≥98%, FG
Sigma-Aldrich
Acetofenona, ReagentPlus®, 99%
Sigma-Aldrich
Butyrophenone, ≥99%
Sigma-Aldrich
Acetofenona, natural, 98%, FG
Sigma-Aldrich
Methyl 4-amino-2-methoxybenzoate, 97%
Sigma-Aldrich
Methyl 4-acetamido-5-chloro-2-methoxybenzoate, 99%