跳转至内容
Merck

GC Analysis of Fresh and Rancid Potato Chips on SPB®-1 SULFUR after SPME using 50/30 μm DVB/Carboxen/PDMS Fiber

GC Analysis of Fresh and Rancid Potato Chips on SPB®-1 SULFUR after SPME using 50/30 μm DVB/Carboxen/PDMS Fiber suitable for GC, application for SPME

材料

分析柱

产品编号
说明
价格

SPB®-1 硫

L × I.D. 30 m × 0.32 mm, df 4.00 μm

固相微萃取纖維

产品编号
说明
价格

SPME 纤维组件二乙烯基苯/羧基/聚二甲基硅氧烷 (DVB/CAR/PDMS)

needle size 24 ga, for use with manual holder

標準

产品编号
说明
价格

癸醛

≥98% (GC), liquid

丁酸甲酯

analytical standard

己酸

analytical standard

辛酸

analytical standard

辛烷

analytical standard

庚醛

≥95%, FCC, FG

壬醛

analytical standard

正己醛

analytical standard

己酸丁酯

analytical standard

辛醛

analytical standard

庚酸

analytical standard

相關產品

产品编号
说明
价格

SPME 光纤支架

for use with manual sampling

SPME 光纤支架

for use with CTC CombiPAL and Gerstel MPS2

CONDITIONS

sample/matrix

3 g crushed potato chips in 15 mL vial

SPME fiber

DVB/Carboxen/PDMS, 50/30 μm, StableFlex (57328-U)

extraction

headspace, 65 °C, 20 min, with stirring

desorption process

3 min, 250 °C

column

SPB-1 SULFUR, 30 m × 0.32 mm I.D., 4.0 μm (24158)

inj. temp.

250 °C

oven

45 °C (1.5 min) to 250 °C at 12 °C/min, hold 10 min

detector

quadrupole mass spectrometer, m/z = 35-290 at 0.6 sec/scan

carrier gas

helium, 40 cm/sec

injection

Splitless/split, closed 2 min

liner

0.75 mm liner

sample

3g crushed potato chips in 15 mL vial

说明

分析報告

Potato chips can contain up to 50% oil and are subject to rancidity. Aldehydes, particularly hexanal and heptanal, are formed from fatty acids that are cleaved at the double bonds when exposed to ultraviolet light and air. A package of 3-year-old potato chips was analyzed using SPME and compared to a fresh package. The presence of aldehydes indicated that the older chips were rancid. However, due to the age of the chips, further oxidation of the aldehydes occurred that resulted in the formation of hexanoic and heptanoic acids. A Carboxen/PDMS/DVB fiber was inserted into the headspace of the potato chip samples at 65 °C for 20 minutes with agitation. The fiber was then analyzed using GC/MS on a SPB-1 SULFUR column. This non-polar column has a thick film and is ideal for smaller molecules.

法律資訊

SPB is a registered trademark of Merck KGaA, Darmstadt, Germany