Skip to Content
Merck
HomePharmacology & Drug Discovery ResearchFamciclovir Tablets USP Monograph Method Using a Purospher™ STAR RP-8 endcapped HPLC Column and UV Detection

Famciclovir Tablets USP Monograph Method Using a Purospher™ STAR RP-8 endcapped HPLC Column and UV Detection

Sonal Shinde, Application Specialist

Article from Analytix Reporter - Issue 6

Introduction

Famciclovir is an antiviral drug indicated for the treatment of herpes zoster, herpes simplex virus 2 (genital herpes), herpes labialis (cold sores), etc. It is a guanosine analogue, a prodrug form of penciclovir, and marketed by Novartis under the trade name Famvir. Generics are produced by TEVA and Mylan, among others. Purospher™ STAR RP-8 endcapped HPLC columns can be used to monitor organic impurities in tablet formulations following the new USP monograph for Famciclovir Tablets. The method suitability requirements are defined by the relative standard deviation (NMT 5.0% for Famciclovir standard solution) and the chromatographic resolution between propionyl famciclovir and 6-chloro famciclovir (NLT 1.2 using the system suitability solution).

Famciclovir

Experimental Conditions

The method acceptance criteria are defined by the relative retention times for Famciclovir related compound A, Famciclovir related compound B, Famciclovir, 6-Chloro famciclovir, and Propionyl famciclovir and are about 0.2, 0.5, 1.0, 1.32, and 1.35, respectively. This application note illustrates with required analytical data that the method meets USP41-NF36 guidelines.

Results

Chromatographic data – blank solution.

Figure 1. Chromatographic data – blank & standard solution.

Chromatographic data –  standard solution.
Chromatographic data - system suitability test (SST) solution.

Figure 2.Chromatographic data - system suitability test (SST) solution.

Chromatographic data - test solution.

Figure 3. Chromatographic data - test solution.

Evaluation and verification

1. Specificity

Inject solution and determine the retention time of desired analyte in the presence of other components such as impurities and excipients.

2. Standard Repeatability (1 ppm)

3.  Linearity, LOD & LOQ

Concentration in ppm

Figure 4.Area / Concentration in ppm

Materials
Loading
Sign In To Continue

To continue reading please sign in or create an account.

Don't Have An Account?