Accéder au contenu
Merck

HPLC Analysis of mAb Charge Variants on TSKgel® CM-STAT

HPLC Analysis of mAb Charge Variants on TSKgel® CM-STAT application for HPLC

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

Colonnes TSKgel® HXL séries GPC

phase carboxymethyl, L × I.D. 10 cm × 4.6 mm, 7 μm particle size

CONDITIONS

column

TSKgel CM-STAT, 10 cm x 4.6 mm I.D., 7 μm particles (821966)

mobile phase

[A] 20 mM MES buffer, pH 6.0; [B] 0.5 M NaCl in buffer A, pH 6.0

gradient

0% B (0 min), 30% B (15 min), 100% B (15 min),

0% B (17 min), 10% B (17 min), 10% B (21 min)

flow rate

1 mL/min

column temp.

ambient

detector

UV, 280 nm

injection

20 μL

sample

monoclonal antibodies (mAb A through E)

Description

Remarque sur l'analyse

A TSKgel CM-STAT weak cation exchange (WCX) column was applied to separate charge variants of several monoclonal antibodies. The typical analysis time on conventional 25 cm long WCX columns of about eighty minutes could be significantly reduced when separation was performed on a 10 cm TSKgel CM-STAT column, filled with 7 μm particles. The analysis profiles for five antibodies show that high resolution analysis can be obtained in about 20 minutes analysis time.

Formule chimique

From "Ion Exchange Chromatography for the Characterization of Biotherapeutics" in The Supelco Reporter, Vol. 29.3, page 20.

Informations légales

TSKgel is a registered trademark of Tosoh Corporation