HPLC Analysis of Aflatoxins on Ascentis® Express C18
![HPLC Analysis of Aflatoxins on Ascentis® Express C18 application for HPLC](/deepweb/assets/sigmaaldrich/product/images/184/506/28bea7ec-8721-4617-a425-f46f92934c37/800/28bea7ec-8721-4617-a425-f46f92934c37.jpg)
Matériaux
Colonne analytique
Réf. du produit
Description
Tarif
CONDITIONS
column
Ascentis Express C18, 10 cm x 2.1 mm I.D., 2.7 μm particles (53823-U)
mobile phase
[A] 5 mM ammonium formate, pH 2.5 (titrated with formic acid); [B] methanol
gradient
5 to 100% B in 10 min
flow rate
0.4 mL/min
pressure
4714 psi (325 bar)
column temp.
35 °C
detector
APCI (+)
injection
5 μL
sample
0.15 - 30 mg/L in 90:10, water:acetonitrile
Description
Remarque sur l'analyse
This application demonstrates the suitability of Ascentis Express C18 for the efficient separation of various toxins.
Informations légales
Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany