HPLC Analysis of 2,2′-Bis[(S)-(4-methylphenyl)sulfinyl]-1,1′-binapthaIene Diastereomers on Ascentis® Express C18
Matériaux
Colonne analytique
Réf. du produit
Description
Tarif
CONDITIONS
column
Ascentis Express C18, 10 cm x 4.6 mm I.D., 2.7 μm particles (53827-U)
mobile phase
[A] water; [B] methanol; (25:75, A:B)
flow rate
0.8 mL/min
pressure
2450 psi (169 bar)
column temp.
35 °C
detector
UV, 240 nm
injection
5 μL
sample
(1R)-2,2′-Bis[(S )-(4-methylphenyl)sulfonyl]-1,1′-binapthalene, 25 μg/mL in methanol; (1S)-2,2′-Bis[(S)-(4-methylphenyl)sulfonyl]-1,1′-binapthalene, 50 μg/mL in methanol
Description
Remarque sur l'analyse
This application demonstrates the suitability of the Ascentis Express C18 for the separation of the diastereomers of by HPLC in reversed phase mode.
Informations légales
Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany