Accéder au contenu
Merck

GC Analysis of Resveratrol in Red Wine on SLB®-5ms after SPME using 85 μm Polyacrylate Fiber

GC Analysis of Resveratrol in Red Wine on SLB®-5ms after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

SLB®-5ms

L × I.D. 30 m × 0.25 mm, df 0.25 μm

Fibre SPME

Réf. du produit
Description
Tarif

Fibres SPME polyacrylate (PA)

df 85 μm(PA, for use with manual holder, needle size 24 ga

Fibres SPME polyacrylate (PA)

df 85 μm(PA, for use with autosampler, needle size 24 ga

Fibres SPME polyacrylate (PA)

df 85 μm(PA, for use with autosampler, needle size 23 ga

Produit(s) apparenté(s)

Réf. du produit
Description
Tarif

Support de fibres SPME

for use with manual sampling

SPME/Direct

pkg of 1 ea

À utiliser avec

Réf. du produit
Description
Tarif

Flacons à vis avec bouchon percé phénolique, pré-assemblés

volume 4 mL, clear glass vial, O.D. × H 15 mm × 45 mm, tan PTFE/silicone septum

Étalon

Réf. du produit
Description
Tarif

Resvératrol

≥99% (HPLC)

CONDITIONS

sample/matrix

3 mL of red wine (California merlot) diluted 3:1 in 12% ethanol:water

SPME fiber

85 μm polyacrylate FS 23 gauge auto (57294-U)

SPME fiber

85 μm polyacrylate FS 24gauge auto (57305)

SPME fiber

85 μm polyacrylate FS 24gauge manual (57304)

extraction

immersion for 15 min at room temperature while stirring at 400 rpm, fiber gently blotted with a Kimwipe to remove excess water, (then derivatization by immersion in a 4 mL vial containing 5 μL of Sylon-BFT (fresh vial for each analysis than was allowed to equilabrate 60-90 minutes prior to use) for 20 min)

desorption process

280 °C for 2 min

column

SLB-5ms; 30 m x 0.25 mm I.D., 0.25 μm (28471-U)

oven

100 °C (1 min), 10 °C/min to 325 °C (3 min)

MSD interface

325 °C

scan range

m/z 40-450

carrier gas

helium, 1 mL/min, constant

liner

0.75 mm I.D., SPME

Description